University Department of Otolaryngology, Head and Neck Surgery, Elfriede-Aulhorn-Straße 5, Tübingen.
MED-EL, Elektromedizinische Geräte GmbH, Starnberg, Germany.
Otol Neurotol. 2019 Dec;40(10):1268-1277. doi: 10.1097/MAO.0000000000002398.
Evaluation of triphasic pulse stimulation in comparison to the traditional biphasic pulse stimulation in cochlear implant (CI) patients with unintended facial nerve costimulation.
Retrospective case review.
Cochlear Implant Center of a University Department of Otolaryngology, Head and Neck Surgery.
Fifteen CI patients (MED-EL, Innsbruck, Austria) received a triphasic fitting map instead of a biphasic fitting map due to a previous diagnosis of facial nerve stimulation or stimulus induced pain during the years 2014 to 2017.
INTERVENTION(S): Application of a triphasic stimulation strategy.
MAIN OUTCOME MEASURE(S): Reduction of facial nerve costimulation and speech understanding. Biphasic and triphasic fitting maps were compared to accurately assess the effects of the switch, and hearing tests (monosyllables and sentences in noise tests) were analyzed.
Triphasic pulse stimulation showed a significant reduction of unintended side effects and resulted in an observed improved quality of life in most cases. Although there was no significant change in the understanding of speech with CI in all test situations, in many cases, improvement was observed.
Triphasic pulse stimulation had a beneficial effect for CI patients with severe, unintended costimulation and should be considered a valuable tool during CI fitting.
评估三脉冲刺激与传统双相脉冲刺激在因意外面神经刺激或刺激诱导疼痛而接受 Cochlear Implant(CI)的患者中的效果。
回顾性病例研究。
耳鼻喉头颈外科大学 Cochlear Implant 中心。
15 名 CI 患者(MED-EL,因斯布鲁克,奥地利)在 2014 年至 2017 年间因先前诊断为面神经刺激或刺激诱导疼痛而接受了三脉冲拟合图,而非双相拟合图。
应用三脉冲刺激策略。
减少面神经刺激和言语理解。比较双相和三相拟合图以准确评估转换效果,并分析听力测试(单音节和噪声测试中的句子)。
三脉冲刺激显著减少了意外的副作用,并在大多数情况下提高了生活质量。尽管在所有测试情况下,CI 患者的言语理解都没有显著改善,但在许多情况下观察到了改善。
三脉冲刺激对因严重意外刺激而接受 Cochlear Implant 的患者有有益的效果,应在 Cochlear Implant 适配过程中被视为一种有价值的工具。