Kristensen Søren Risom, Nybo Jette, Pedersen Shona
Department of Clinical Biochemistry Aalborg University Hospital Aalborg Denmark.
Department of Clinical Medicine Aalborg University Aalborg Denmark.
Res Pract Thromb Haemost. 2022 Jan 31;6(1):e12654. doi: 10.1002/rth2.12654. eCollection 2022 Jan.
The thrombin generation (TG) assay, which measures global coagulation, has mainly been used as a research tool to investigate thrombotic and bleeding disorders. Recently, Diagnostica Stago launched the ST Genesia, a fully automated system to perform "routine version" of this assay. The objectives of this study were to evaluate the imprecision compared with the previous method, Thrombinoscope CAT, and to establish reference intervals.
Thrombin generation was measured in platelet-poor citrated plasma from 20 normal controls (fresh and after freezing at -80°C up to 12-13 weeks) on CAT and ST Genesia in duplicate to estimate the total variation, and within and between variations. The reference intervals were estimated nonparametrically in 30 men, 30 women taking combined oral contraceptives (COCs), and 30 women not taking COCs. These were sampled in both Vacutainer and Monovette tubes (i.e., tubes with a high and minimal contact activation, respectively).
Freezing had minimal effects. Imprecision was comparable between the ST Genesia and CAT, with a strong correlation between the two methods. TG was higher when sampled in Vacutainer than in Monovette. We observed a distinct difference between women taking and not taking COCs, whereas men and women not taking COC were quite similar.
Thrombin generation on ST Genesia showed an analytical variation similar to that of CAT. The results depended on the type of sample tubes; thus, reference intervals must be established for the collection tubes used in each laboratory. Furthermore, a considerable difference was observed between women using and not using COCs.
凝血酶生成(TG)检测用于测量整体凝血功能,主要作为研究血栓形成和出血性疾病的工具。最近,Diagnostica Stago公司推出了ST Genesia,这是一种用于进行该检测“常规版本”的全自动系统。本研究的目的是评估与之前的Thrombinoscope CAT方法相比的不精密度,并建立参考区间。
在20名正常对照者(新鲜样本以及在-80°C冷冻长达12 - 13周后的样本)的乏血小板枸橼酸盐血浆中,使用CAT和ST Genesia重复测量凝血酶生成,以估计总变异、批内变异和批间变异。对30名男性、30名服用复方口服避孕药(COC)的女性和30名未服用COC的女性非参数估计参考区间。这些样本分别采集于真空管(Vacutainer)和单管采血管(Monovette)中(即分别具有高接触激活和最小接触激活的采血管)。
冷冻的影响极小。ST Genesia和CAT之间的不精密度相当,两种方法之间具有很强的相关性。在真空管中采样时的TG高于在单管采血管中采样时的TG。我们观察到服用和未服用COC的女性之间存在明显差异,而未服用COC的男性和女性则非常相似。
ST Genesia上的凝血酶生成显示出与CAT相似的分析变异。结果取决于采血管的类型;因此,每个实验室必须为所使用的采血管建立参考区间。此外,观察到使用和未使用COC的女性之间存在显著差异。