Department of Medical Physics, Guy's and St Thomas' Hospitals NHS Trust, London, UK.
One Small Step Gait Laboratory, Guy's and St Thomas' Hospitals NHS Trust, London, UK.
BMJ Open Respir Res. 2022 Feb;9(1). doi: 10.1136/bmjresp-2021-001159.
During the COVID-19 pandemic, portable pulse oximeters were issued to some patients to permit home monitoring and alleviate pressure on inpatient wards. Concerns were raised about the accuracy of these devices in some patient groups. This study was conducted in response to these concerns.
To evaluate the performance characteristics of five portable pulse oximeters and their suitability for deployment on home-use pulse oximetry pathways created during the COVID-19 pandemic. This study considered the effects of different device models and patient characteristics on pulse oximeter accuracy, false negative and false positive rate.
A total of 915 oxygen saturation (sO) measurements, paired with measurements from a hospital-standard pulse oximeter, were taken from 50 patients recruited from respiratory wards and the intensive care unit at an acute hospital in London. The effects of device model and several patient characteristics on bias, false negative and false positive likelihood were evaluated using multiple regression analyses.
All five portable pulse oximeters appeared to outperform the standard to which they were manufactured. Device model, patient sO and patient skin colour were significant predictors of measurement bias, false positive and false negative rate, with some variation between models. The false positive and false negative rates were 11.2% and 24.5%, respectively, with substantial variation between models.
在 COVID-19 大流行期间,为了允许患者在家中进行监测并减轻住院病房的压力,向一些患者发放了便携式脉搏血氧仪。人们对这些设备在某些患者群体中的准确性表示担忧。这项研究正是针对这些担忧开展的。
评估五款便携式脉搏血氧仪的性能特征及其在 COVID-19 大流行期间创建的家庭使用脉搏血氧仪路径中的适用性。本研究考虑了不同设备型号和患者特征对脉搏血氧仪准确性、假阴性和假阳性率的影响。
从伦敦一家急性医院的呼吸病房和重症监护病房招募了 50 名患者,对其进行了 915 次血氧饱和度(sO)测量,同时与医院标准脉搏血氧仪的测量结果进行了配对。使用多元回归分析评估了设备型号和多个患者特征对偏差、假阴性和假阳性似然率的影响。
所有五款便携式脉搏血氧仪似乎都优于其制造标准。设备型号、患者的 sO 和患者的皮肤颜色是测量偏差、假阳性和假阴性率的显著预测因素,不同型号之间存在一定差异。假阳性和假阴性率分别为 11.2%和 24.5%,不同型号之间存在较大差异。