Suppr超能文献

前置场域研究:当使用 SITA-Faster 时,假阳性和起始点误差对视场可靠性的影响。

The Frontloading Fields Study: The Impact of False Positives and Seeding Point Errors on Visual Field Reliability When Using SITA-Faster.

机构信息

Centre for Eye Health, University of New South Wales, Kensington, New South Wales, Australia.

School of Optometry and Vision Science, University of New South Wales, Kensington, New South Wales, Australia.

出版信息

Transl Vis Sci Technol. 2022 Feb 1;11(2):20. doi: 10.1167/tvst.11.2.20.

Abstract

PURPOSE

The purpose of this study was to evaluate the impact of two conventional reliability criteria (false positives [FPs] and seeding point errors [SPEs]) and the concurrent effect of low sensitivity points (≤19 dB) on intrasession SITA-Faster visual field (VF) result correlations.

METHODS

There were 2320 intrasession SITA-Faster VF results from 1160 eyes of healthy, glaucoma suspects, and subjects with glaucoma that were separated into "both reliable" or "reliable-unreliable" pairs. VF results (mean deviation and pointwise sensitivity) were analyzed against the spectrum of FP rates and SPE, with and without censorship of sensitivity results ≤19 dB. Segmental linear regression was used to identify critical points where visual field results were significantly different between tests due to FP levels.

RESULTS

There was a significant, but small (0.09 dB per 1% exceeding 12%) increase in mean deviation, and an increase in the number of points showing a >3 dB sensitivity increase (0.25-0.28 locations per 1% exceeding 12%). SPEs were almost exclusively related to a decrease in sensitivity at the primary seeding points but did not result in significant differences in other indices. Censoring sensitivity results ≤19 dB significantly improved the correlation between reliable and unreliable results.

CONCLUSIONS

Current criteria for judging an unreliable VF result (FP rate >15% and SPE) can lead to data being erroneously excluded, as many results do not show significant differences compared to those deemed "reliable." Censoring of sensitivity results ≤19 dB improves intrasession correlations in VF results.

TRANSLATIONAL RELEVANCE

We provide guidelines for assessing the impact of FP, SPE, and low sensitivity results on VF interpretation.

摘要

目的

本研究旨在评估两种传统可靠性标准(假阳性[FP]和起始点误差[SPE])的影响,以及低灵敏度点(≤19dB)对单次 SITA-Faster 视野(VF)结果相关性的并发影响。

方法

从 1160 只健康、疑似青光眼和青光眼患者的眼中,共获得 2320 对单次 SITA-Faster VF 结果,将其分为“均可靠”或“可靠-不可靠”对。VF 结果(平均偏差和逐点灵敏度)与 FP 率和 SPE 的范围进行了分析,同时对灵敏度结果≤19dB 进行了屏蔽。采用分段线性回归来确定由于 FP 水平导致两次测试之间视野结果显著不同的临界点。

结果

平均偏差有显著但较小的增加(每超过 12%增加 0.09dB),敏感性增加>3dB 的点数也有所增加(每超过 12%增加 0.25-0.28 个位置)。SPE 几乎完全与主要起始点的敏感性降低有关,但不会导致其他指标出现显著差异。对灵敏度结果≤19dB 进行屏蔽,可显著提高可靠和不可靠结果之间的相关性。

结论

当前用于判断不可靠 VF 结果的标准(FP 率>15%和 SPE)可能导致数据被错误排除,因为许多结果与被认为“可靠”的结果相比并没有显著差异。对灵敏度结果≤19dB 的屏蔽可改善 VF 结果的单次内相关性。