Piotrowski Michał, Maras Piotr, Kadłubowski Sławomir, Kozicki Marek
Department of Mechanical Engineering, Informatics and Chemistry of Polymer Materials, Faculty of Materials Technologies and Textile Design, Lodz University of Technology, 90-543 Lodz, Poland.
Department of Radiotherapy Planning, Copernicus Hospital, 93-513 Lodz, Poland.
Materials (Basel). 2022 Jan 27;15(3):984. doi: 10.3390/ma15030984.
This paper presents the results of research on the Fricke-XO-Pluronic F-127 dosimeter. It consists of a Fricke dosimetric solution and xylenol orange (XO), which are embedded in a matrix of poly(ethylene oxide)--poly(propylene oxide)--poly(ethylene oxide) (Pluronic F-127). Upon irradiation, Fe ions transform into Fe, forming a colored complex with XO ([XO-Fe]). The color intensity is related to the dose absorbed. The optimal composition, storage conditions, and radiation-induced performance of the Fricke-XO-Pluronic F-127 dosimeter were investigated. The optimal composition was found to be 1 mM FAS, 50 mM sulfuric acid (HSO), 0.165 mM XO in 25% Pluronic F-127. The basic features of this dosimeter are discussed, such as dose sensitivity, linear and dynamic dose range, stability before and after irradiation, storage conditions, dose response for irradiation with 6 and 15 MV photons, and batch-to-batch reproducibility. The obtained results showed a certain potential of the Fricke-XO-Pluronic F-127 for radiotherapy dosimetry.
本文介绍了对弗里克-XO-普朗尼克F-127剂量计的研究结果。它由弗里克剂量测定溶液和二甲酚橙(XO)组成,二者嵌入聚(环氧乙烷)-聚(环氧丙烷)-聚(环氧乙烷)(普朗尼克F-127)的基质中。辐照时,铁离子转变为铁,与XO形成有色络合物([XO-Fe])。颜色强度与吸收剂量相关。研究了弗里克-XO-普朗尼克F-127剂量计的最佳组成、储存条件和辐射诱导性能。发现最佳组成为25%普朗尼克F-127中含1 mM FAS、50 mM硫酸(HSO)、0.165 mM XO。讨论了该剂量计的基本特性,如剂量敏感性、线性和动态剂量范围、辐照前后的稳定性、储存条件、6和15 MV光子辐照的剂量响应以及批次间的重现性。所得结果表明弗里克-XO-普朗尼克F-127在放射治疗剂量测定方面具有一定潜力。