Suppr超能文献

放射性核素心肌灌注显像定量分析:Carimas 软件在整体、局部和节段水平的观察者间一致性良好。

Myocardial perfusion quantification with Rb-82 PET: good interobserver agreement of Carimas software on global, regional, and segmental levels.

机构信息

Turku PET Centre, University of Turku and Turku University Hospital, Turku, Finland.

IM Sechenov Institute of Evolutionary Physiology and Biochemistry RAS, Saint Petersburg, Russia.

出版信息

Ann Nucl Med. 2022 Jun;36(6):507-514. doi: 10.1007/s12149-022-01729-5. Epub 2022 Feb 22.

Abstract

PURPOSE

To estimate the interobserver agreement of the Carimas software package (SP) on global, regional, and segmental levels for the most widely used myocardial perfusion PET tracer-Rb-82.

MATERIALS AND METHODS

Rest and stress Rb-82 PET scans of 48 patients with suspected or known coronary artery disease (CAD) were analyzed in four centers using the Carimas SP. We considered values to agree if they simultaneously had an intraclass correlation coefficient (ICC) > 0.75 and a difference < 20% of the median across all observers.

RESULTS

The median values on the segmental level were 1.08 mL/min/g for rest myocardial blood flow (MBF), 2.24 mL/min/g for stress MBF, and 2.17 for myocardial flow reserve (MFR). For the rest MBF and MFR, all the values at all the levels fulfilled were in excellent agreement. For stress MBF, at the global and regional levels, all the 24 comparisons showed excellent agreement. Only 1 out of 102 segmental comparisons (seg. 14) was over the adequate agreement limit-23.5% of the median value (ICC = 0.95).

CONCLUSION

Interobserver agreement for Rb-82 PET myocardial perfusion quantification analyzed with Carimas is good at any LV segmentation level-global, regional, and segmental. It is good for all the estimates-rest MBF, stress MBF, and MFR.

摘要

目的

评估 Carimas 软件包(SP)在全球、区域和节段水平上对最广泛使用的心肌灌注 PET 示踪剂-Rb-82 的观察者间一致性。

材料和方法

对 48 例疑似或已知冠心病(CAD)患者的静息和应激 Rb-82 PET 扫描使用 Carimas SP 在四个中心进行分析。我们认为,如果所有观察者同时具有大于 0.75 的组内相关系数(ICC)和小于中位数 20%的差异,则认为值是一致的。

结果

节段水平的中位数为静息心肌血流(MBF)为 1.08 mL/min/g,应激 MBF 为 2.24 mL/min/g,心肌血流储备(MFR)为 2.17。对于静息 MBF 和 MFR,所有水平的所有值都具有极好的一致性。对于应激 MBF,在全球和区域水平上,所有 24 次比较均显示出极好的一致性。只有 1 次节段比较(节段 14)超过了足够的一致性限制-中位数的 23.5%(ICC=0.95)。

结论

使用 Carimas 分析的 Rb-82 PET 心肌灌注定量的观察者间一致性在任何 LV 分段水平-全局、区域和节段均良好。对于所有估计值-静息 MBF、应激 MBF 和 MFR 都是如此。