分析在一家牙科机构中,锥形束计算机断层扫描(CBCT)的患者剂量与扫描设置和成像指征之间的关系,以建立机构诊断参考水平。

An analysis of patient dose received during cone-beam computed tomography in relation to scan settings and imaging indications as seen in a dental institution in order to establish institutional diagnostic reference levels.

机构信息

Oral and Maxillofacial Radiology, Applied Oral Sciences and Community Dental Care, Faculty of Dentistry, The University of Hong Kong, Hong Kong SAR, China.

Oral and Maxillofacial Surgery, Faculty of Dentistry, The University of Hong Kong, Hong Kong SAR, China.

出版信息

Dentomaxillofac Radiol. 2022 Jul 1;51(5):20200529. doi: 10.1259/dmfr.20200529. Epub 2022 Mar 21.

Abstract

OBJECTIVES

To investigate the dose-area product (DAP) of cone-beam computed tomography (CBCT) examinations for different scan settings and imaging indications, and to establish institutional diagnostic reference levels (DRLs) for dose optimisation.

METHODS

A retrospective analysis of the DAP values of 3568 CBCT examinations taken from two different devices at the Prince Philip Dental Hospital, Hong Kong between 2016 and 2021 was performed. Patient- (age, gender, and imaging indication) and imaging-related (CBCT device, field-of-view (FOV), and voxel size) were correlated with the DAPs. The indication-oriented third-quartile DAP values were compared with DRLs from the UK, Finland, and Switzerland. The obtained third-quartile DAPs lower than the national DRLs and those for which no national DRLs have been proposed were used to establish institutional DRLs.

RESULTS

In the investigated CBCTs, the DAP value for large FOV scans was significantly lower than medium/small FOVs. CBCTs with a small voxel size exhibited a significantly higher DAP than those with a medium/large voxel size. CBCTs for endodontic, periodontal, orthodontic, or orthognathic evaluation exhibited a significantly higher DAP than other indications. Twelve indication-oriented institutional DRLs were established and five of them were lower than the national DRLs: third molars (229 mGy×cm), jaw cysts/tumours (410 mGy×cm), maxillary sinus pathology (520 mGy×cm), developing dentition (164 mGy×cm), and periapical lesions (564 mGy×cm).

CONCLUSIONS

CBCT examinations for endodontic, periodontal, orthodontic, or orthognathic evaluation may deliver a higher radiation dose to the patient than other imaging tasks. A periodic review of the patient dose from CBCT imaging and establishment of institutional DRLs for specific clinical settings are needed for monitoring patient dose and to optimise indication-oriented scanning protocols.