TonoVet Plus®: Higher reliability and repeatability compared with Tono-Pen XL™ and TonoVet® in rabbits.

机构信息

Veterinary Medicine and Zootechny, Federal University of Bahia, Salvador, Bahia, Brazil.

出版信息

Vet Ophthalmol. 2022 Jul;25(4):272-281. doi: 10.1111/vop.12981. Epub 2022 Mar 5.

Abstract

OBJECTIVE

To evaluate the accuracy and repeatability of the Tono-Pen XL™, TonoVet® and TonoVet Plus® tonometers by manometric evaluation, and to establish adjustment equations for intraocular pressure (IOP) estimates in rabbits.

ANIMAL STUDIED

Rabbits.

PROCEDURES

A postmortem study was conducted on seven rabbit eyes to verify the correlation between manometry and tonometry with an artificial incremental increase in IOP from 5 and 60 mmHg. A clinical study was conducted to evaluate accuracy and to establish reference values for the species, with measurement of IOP in 17 animals, for 2 consecutive days, with the same tonometers and calibrations used in the postmortem evaluations.

RESULTS

There were strong linear trends for all evaluated tonometers. In the in-vivo evaluation, the mean IOP values were: 14.23 ± 1.75 (Tono-Pen XL™); 13.89 ± 2.07 (TonoVet® calibration mode 'd'); 8.88 ± 1.24 (TonoVet calibration mode 'p'); 18.59 ± 1.94 (Tonovet Plus®). There was a significant difference in the two evaluation times for the two TonoVet® calibration modes. The adjustment equations generated from the manometry for the evaluated tonometers were: Y = 0.2570X + 2.219 (Tono-Pen XL™), Y = 0.2289X + 2.389 (TonoVet® 'd'), Y = 0.4043X + 4.062 (TonoVet® 'p'), Y = 0.1233X + 0.3644 (TonoVet Plus®) (X is device-estimated IOP).

CONCLUSIONS

All evaluated tonometers were well correlated with the manometry, with an underestimation of IOP by all devices. Applying adjustment formulas may compensate for systematic errors. TonoVet Plus® was well tolerated, and showed better repeatability and reliability in successive evaluations.

摘要

目的

通过眼压计的压力测量评估 Tono-Pen XL™、TonoVet®和 TonoVet Plus®眼压计的准确性和可重复性,并建立兔眼眼压(IOP)估计值的调整方程。

研究对象

兔。

操作流程

对七只兔眼进行尸体研究,以验证眼压计和眼压计之间的相关性,方法是人工递增眼压从 5 至 60mmHg。进行临床研究以评估准确性,并为该物种建立参考值,在 17 只动物中连续两天用相同的眼压计和相同的校准值进行测量。

结果

所有评估的眼压计都呈现出强烈的线性趋势。在体内评估中,平均 IOP 值为:14.23±1.75mmHg(Tono-Pen XL™);13.89±2.07mmHg(TonoVet®校准模式“d”);8.88±1.24mmHg(TonoVet 校准模式“p”);18.59±1.94mmHg(Tonovet Plus®)。两种 TonoVet®校准模式的两次评估之间存在显著差异。从压力测量生成的评估眼压计的调整方程为:Y=0.2570X+2.219(Tono-Pen XL™),Y=0.2289X+2.389(TonoVet®“d”),Y=0.4043X+4.062(TonoVet®“p”),Y=0.1233X+0.3644(TonoVet Plus®)(X 为设备估计的IOP)。

结论

所有评估的眼压计均与眼压计高度相关,所有设备均低估了 IOP。应用调整公式可能会补偿系统误差。TonoVet Plus®耐受性良好,在连续评估中表现出更好的可重复性和可靠性。