Evaluation of the effect of disposable tonometer cover brand on performance of Tono-Pen Vet in canine eyes.

机构信息

Department of Surgical Sciences, School of Veterinary Medicine, University of Wisconsin-Madison, Madison, WI, USA.

Department of Biostatistics and Medical Informatics, Institute for Clinical and Translational Research (ICTR, University of Wisconsin-Madison, Madison, WI, USA.

出版信息

Vet Ophthalmol. 2021 Mar;24 Suppl 1(Suppl 1):194-198. doi: 10.1111/vop.12877. Epub 2021 Feb 27.

Abstract

PRIMARY OBJECTIVE

To evaluate the effect of latex tip cover manufacturer on accuracy and repeatability of Tono-Pen Vet™ in canine eyes.

ANIMAL STUDIED

Twelve enucleated globes from six dogs.

PROCEDURES

The anterior chamber was cannulated and connected to a calibrated manometer. Intraocular pressure (IOP) measurements were obtained using the Tono-Pen Vet and TONOVET Plus at manometric IOP ranging from 5 to 80 mmHg. At each IOP, the Tono-Pen Vet was used with a new Ocu-Film™ latex tip cover (the only manufacturer-approved brand of cover) followed by a new Softips™ latex tip cover. For comparison, the TONOVET Plus was also used at each IOP with a new disposable rebound probe. Measured IOP values were analyzed by linear regression and intraclass correlation coefficient (ICC).

RESULTS

Tono-Pen Vet accuracy was unaffected by tip cover manufacturer or by frequent change in cover. Using ICC analysis, repeatability of measurements using either tonometer was good to excellent at physiologic IOP levels but variably decreased with both devices at supraphysiologic IOP.

CONCLUSIONS

Neither tip cover manufacturer nor frequent changes in tip cover adversely affect Tono-Pen Vet accuracy. Measurement repeatability with Tono-Pen Vet and TONOVET Plus is widely variable at supraphysiologic IOP. Therefore, minor changes in IOP >25 mmHg should not be used to make clinical decisions without considering this variability.

摘要

主要目标

评估乳胶探头盖制造商对犬眼 Tono-Pen Vet™准确性和可重复性的影响。

研究动物

来自 6 只狗的 12 个眼球。

程序

前房被穿刺并连接到校准的压力计。使用 Tono-Pen Vet 和 TONOVET Plus 在眼压(IOP)范围为 5 至 80mmHg 的情况下进行眼压测量。在每个 IOP 下,使用新的 Ocu-Film™乳胶探头盖(唯一经制造商批准的探头盖品牌)和新的 Softips™乳胶探头盖后,使用 Tono-Pen Vet 进行测量。为了比较,在每个 IOP 下,还使用新的一次性回弹探头对 TONOVET Plus 进行测量。分析测量的眼压值采用线性回归和组内相关系数(ICC)。

结果

探头盖制造商或探头盖的频繁更换均不影响 Tono-Pen Vet 的准确性。使用 ICC 分析,两种眼压计在生理 IOP 水平下的测量重复性均为良好至优秀,但在两种设备的高眼压下,重复性均有所下降。

结论

探头盖制造商或探头盖的频繁更换均不会对 Tono-Pen Vet 的准确性产生不利影响。Tono-Pen Vet 和 TONOVET Plus 的测量重复性在高眼压时差异较大。因此,在不考虑这种变异性的情况下,不应仅根据 IOP 变化>25mmHg 来做出临床决策。

相似文献

1
Evaluation of the effect of disposable tonometer cover brand on performance of Tono-Pen Vet in canine eyes.
Vet Ophthalmol. 2021 Mar;24 Suppl 1(Suppl 1):194-198. doi: 10.1111/vop.12877. Epub 2021 Feb 27.
2
Validation and comparison of four handheld tonometers in normal ex vivo canine eyes.
Vet Ophthalmol. 2021 Mar;24 Suppl 1(Suppl 1):162-170. doi: 10.1111/vop.12780. Epub 2020 Jun 1.
4
Tonometer validation and intraocular pressure reference values in the normal chinchilla (Chinchilla lanigera).
Vet Ophthalmol. 2018 Jan;21(1):4-9. doi: 10.1111/vop.12468. Epub 2017 Mar 17.
5
Calibration of the TonoVet and Tono-Pen Vet tonometers in the porcine eye.
Vet Ophthalmol. 2017 Nov;20(6):571-573. doi: 10.1111/vop.12445. Epub 2016 Nov 24.
7
Comparison among TonoVet, TonoVet Plus, Tono-Pen Avia Vet, and Kowa HA-2 portable tonometers for measuring intraocular pressure in dogs.
Vet World. 2021 Sep;14(9):2444-2451. doi: 10.14202/vetworld.2021.2444-2451. Epub 2021 Sep 21.
8
Validation of the TonoVet® rebound tonometer in normal and glaucomatous cats.
Vet Ophthalmol. 2013 Mar;16(2):111-8. doi: 10.1111/j.1463-5224.2012.01038.x. Epub 2012 Jun 6.
9
Validation of the Reichert® Tono-Vera® Vet rebound tonometer in normal ex vivo canine eyes.
Vet Ophthalmol. 2024 May;27(3):290-293. doi: 10.1111/vop.13213. Epub 2024 Apr 4.
10
Validation of the Icare TONOVET plus rebound tonometer in normal rabbit eyes.
Exp Eye Res. 2019 Aug;185:107698. doi: 10.1016/j.exer.2019.107698. Epub 2019 Jun 12.

本文引用的文献

1
Validation and comparison of four handheld tonometers in normal ex vivo canine eyes.
Vet Ophthalmol. 2021 Mar;24 Suppl 1(Suppl 1):162-170. doi: 10.1111/vop.12780. Epub 2020 Jun 1.
2
Validation of the Icare TONOVET plus rebound tonometer in normal rabbit eyes.
Exp Eye Res. 2019 Aug;185:107698. doi: 10.1016/j.exer.2019.107698. Epub 2019 Jun 12.
3
Evaluation of three hand-held tonometers in normal canine eyes.
Vet J. 2017 Jun;224:7-10. doi: 10.1016/j.tvjl.2017.04.003. Epub 2017 May 2.
4
Calibration of the TonoVet and Tono-Pen Vet tonometers in the porcine eye.
Vet Ophthalmol. 2017 Nov;20(6):571-573. doi: 10.1111/vop.12445. Epub 2016 Nov 24.
6
Validation of the TonoVet® rebound tonometer in normal and glaucomatous cats.
Vet Ophthalmol. 2013 Mar;16(2):111-8. doi: 10.1111/j.1463-5224.2012.01038.x. Epub 2012 Jun 6.
7
Intraclass correlations: uses in assessing rater reliability.
Psychol Bull. 1979 Mar;86(2):420-8. doi: 10.1037//0033-2909.86.2.420.
8
Comparison of the use of new handheld tonometers and established applanation tonometers in dogs.
Am J Vet Res. 2006 Jan;67(1):134-44. doi: 10.2460/ajvr.67.1.134.
9
Evaluation of a rebound tonometer for measuring intraocular pressure in dogs and horses.
J Am Vet Med Assoc. 2005 Jul 15;227(2):244-8. doi: 10.2460/javma.2005.227.244.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验