Saghir Nabeel, Aggarwal Arjun, Soneji Nisha, Valencia Victoria, Rodgers George, Kurian Thomas
Dell Medical School at The University of Texas at Austin, Austin, Texas - Department of Internal Medicine.
Ascension Seton Medical Center, Austin, Texas.
Cardiovasc Digit Health J. 2020 Aug 26;1(1):30-36. doi: 10.1016/j.cvdhj.2020.07.002. eCollection 2020 Jul-Aug.
The Apple Watch Series 4 (AW) can detect atrial fibrillation and perform a single-lead electrocardiogram (ECG), but the clinical accuracy of AW ECG waveforms compared to lead 1 of a 12-lead ECG is unclear.
The purpose of this study was to assess the accuracy of interval measurements on AW ECG tracings in comparison to lead 1 on a 12-lead ECG.
We obtained ECGs at a university hospital of healthy volunteers age >18 years. ECG waveforms were measured with calipers to the nearest 0.25 mm. When possible, 3 consecutive waveforms in lead 1 were measured. Waveform properties, including intervals, were recorded. Concordance correlation coefficients and Bland-Altman plots were used to assess level of agreement between devices.
Twelve-lead (n = 113) and AW (n = 129) ECG waveforms from 43 volunteers (mean age 31 years; 65% female) were analyzed. Sinus rhythm interpretation between devices was 100% concordant. No arrhythmias were recorded. Mean difference (d) for heart rate was 1.16 ± 4.33 bpm (r = 0.94); 3.83 ± 113.54 ms for RR interval (r = 0.79); 5.43 ± 17 ms for PR interval (r = 0.83); -6.89 ± 14.81 ms for QRS interval (r = 0.65); -11.27 ± 22.9 ms for QT interval (r = 0.79); and -11.67 ± 27 ms for QTc interval (r = 0.57). There was moderate (d <40 ms) to strong (d <20 ms or < 5 bpm) agreement between devices represented by Bland-Altman plots.
The AW produces accurate ECGs in healthy adults with moderate to strong agreement of basic ECG intervals.
苹果手表系列4(AW)能够检测心房颤动并进行单导联心电图(ECG)检查,但与12导联心电图的第1导联相比,AW心电图波形的临床准确性尚不清楚。
本研究旨在评估与12导联心电图的第1导联相比,AW心电图描记图上间期测量的准确性。
我们在一家大学医院获取了年龄大于18岁的健康志愿者的心电图。使用卡尺测量心电图波形,精确到最接近的0.25毫米。尽可能测量第1导联中连续的3个波形。记录包括间期在内的波形属性。使用一致性相关系数和布兰德-奥特曼图来评估不同设备之间的一致程度。
分析了43名志愿者(平均年龄31岁;65%为女性)的12导联(n = 113)和AW(n = 129)心电图波形。不同设备之间窦性心律的解读完全一致。未记录到心律失常。心率的平均差值(d)为1.16±4.33次/分钟(r = 0.94);RR间期为3.83±113.54毫秒(r = 0.79);PR间期为5.43±17毫秒(r = 0.83);QRS间期为-6.89±14.81毫秒(r = 0.65);QT间期为-11.27±22.9毫秒(r = 0.79);QTc间期为-11.67±27毫秒(r = 0.57)。布兰德-奥特曼图显示不同设备之间存在中度(d <40毫秒)到高度(d <20毫秒或<5次/分钟)的一致性。
AW在健康成年人中可产生准确的心电图,基本心电图间期具有中度到高度的一致性。