Griffiths C J, Murray A, Ramsden P D
J Urol. 1986 Oct;136(4):808-12. doi: 10.1016/s0022-5347(17)45086-5.
We assessed 6 methods for calculating bladder volume from ultrasonic cross-sectional scans. The technique that used the largest number of features from the scans gave the best results. For volumes greater than 150 ml. an accuracy of 0.87 and repeatability of +/- 9 per cent (standard deviation) were obtained.