Int J Environ Res Public Health. 2022 Apr 13;19(8):4719. doi: 10.3390/ijerph19084719.
Background: We compare the accuracy of new intraoral scanners (IOSs) in full-arch digital implant impressions. Methods: A master model with six scan bodies was milled in poly(methyl methacrylate), measured by using a coordinate measuring machine, and scanned 15 times with four IOSs: PrimeScan, Medit i500, Vatech EZ scan, and iTero. The software was developed to identify the position points on each scan body. The 3D position and distance analysis were performed. Results: The average and ± standard deviation of the 3D position analysis was 29 μm ± 6 μm for PrimeScan, 39 μm ± 6 μm for iTero, 48 μm ± 18 μm for Mediti500, and 118 μm ± 24 μm for Vatech EZ scan (p < 0.05). Conclusions: All IOSs are able to make a digital complete implant impression in vitro according to the average misfit value reported in literature (150 μm); however, the 3D distance analysis showed that only the Primescan and iTero presented negligible systematic error sources.
我们比较了几种新型口内扫描仪(IOS)在全口数字化种植体印模中的准确性。
通过使用坐标测量机对一个带有六个扫描体的母模进行铣削,并用四种 IOS(PrimeScan、Medit i500、Vatech EZ scan 和 iTero)扫描 15 次。开发了一种软件来识别每个扫描体上的位置点。进行三维位置和距离分析。
三维位置分析的平均值和标准偏差分别为 PrimeScan 为 29μm±6μm,iTero 为 39μm±6μm,Medit i500 为 48μm±18μm,Vatech EZ scan 为 118μm±24μm(p<0.05)。
根据文献报道的平均(150μm)不匹配值,所有 IOS 都能够在体外进行数字化完整种植体印模,但三维距离分析表明只有 PrimeScan 和 iTero 显示出可忽略的系统误差源。