评价七种不同的口内数字化印模系统和一台实验室扫描仪在人类上颌骨上获取全牙弓数字化印模的真实性和精确性。

Evaluation of the trueness and precision of complete arch digital impressions on a human maxilla using seven different intraoral digital impression systems and a laboratory scanner.

机构信息

Department of Oral Rehabilitation, Medical University of South Carolina College of Dental Medicine, Charleston, South Carolina.

Department of Stomatology, Medical University of South Carolina College of Dental Medicine, Charleston, South Carolina.

出版信息

J Esthet Restor Dent. 2019 Jul;31(4):369-377. doi: 10.1111/jerd.12485. Epub 2019 May 6.

Abstract

OBJECTIVES

An impression accuracy study using a cadaver maxilla was performed using both prepared and intact teeth as well as palatal tissue.

MATERIALS AND METHODS

Three crown preparations were performed on a cadaver maxilla. Seven different digital impression systems along with polyvinylsiloxane impressions were used to create digital models of the maxilla. Three-dimensional (3D) files of the experimental models were compared to a master model. The 3D files were overlaid and analyzed using a comparison software to create color coded figures that were measured for deviations between the master and experimental models.

RESULTS

For scanning tooth structure, only the Planscan was significantly less accurate than the rest of impression techniques. No significant differences in accuracy were found between models created using digital impressions and those created from traditional vinyl polysiloxane impressions with cross arch deviations ranging from 18 to 39 μm for each.

CONCLUSIONS

Impressions taken using all digital impression systems, save for the Planscan, were able to accurately replicate the tissues of a complete arch human maxilla.

CLINICAL SIGNIFICANCE

Studies examining accuracy of digital impression systems have generally been performed on materials other than dental tissues. Optically, materials such as plastic and metal have properties different from enamel and dentin. This study evaluates accuracy of digital impression systems on human dentin, enamel, and soft tissues.

摘要

目的

使用尸体上颌骨进行了印模精度研究,同时使用预备牙和完整牙以及腭组织。

材料和方法

在尸体上颌骨上进行了 3 个牙冠预备。使用 7 种不同的数字化印模系统和聚硅氧烷印模来创建上颌骨的数字模型。将实验模型的三维 (3D) 文件与主模型进行比较。使用比较软件覆盖和分析 3D 文件,以创建颜色编码的图形,然后测量主模型和实验模型之间的偏差。

结果

对于扫描牙结构,仅 Planscan 明显不如其他印模技术准确。使用数字印模创建的模型与使用传统乙烯基聚硅氧烷制作的模型之间没有发现准确性差异,每个模型的跨弓偏差范围为 18 至 39μm。

结论

除了 Planscan 之外,所有数字印模系统的印模都能够准确复制完整上颌牙弓的组织。

临床意义

研究数字印模系统准确性的研究通常是在除了牙科组织之外的其他材料上进行的。从光学上讲,塑料和金属等材料的特性与牙釉质和牙本质不同。本研究评估了数字印模系统在人牙本质、牙釉质和软组织上的准确性。