Abstract

Glucometer is the most commonly used POCT device and guides monitoring of blood glucose level in both clinical settings and outside. Inaccurate glucometer readings resulting in erroneous therapeutic intervention has critical consequences on patient care. Regulatory guidelines for performance evaluation of glucometers are not available in many countries. A robust program implemented by the hospital is essential to ensure accuracy and precision of glucometers to produce optimal results. The objective of this study was to design a quality assurance program for the evaluation of glucometers in a high volume tertiary care referral hospital and evaluate the results from July'18 to July'19. Seventy three glucometers used across the hospital were subjected to Internal Quality Control checks and Proficiency Testing performed once a month and every 3 months respectively. The results were reviewed and plotted on a Bland Altman Graph. Clarke Error Grid Analysis was done to evaluate the clinical significance of inaccuracies in the measurement of blood glucose concentration as per ISO 15197: 2013. Eight devices were identified as unacceptable by ISO standards and replaced subsequently. 96.83% and 3.17% of the values were in Zone A and B of Clarke Error Grid Analysis. The study complied with the standard which requires that 99% of the values fall within zones A and B. The review of the program after one year and its ability to identify defective glucometers has validated the efficacy of the model. The method used may be suggested as a prototype for quality management of glucometers in a clinical setting.

摘要

血糖仪是最常用的即时检验设备,可指导临床环境内外的血糖水平监测。血糖仪读数不准确导致错误的治疗干预会对患者护理产生严重后果。许多国家没有血糖仪性能评估的监管指南。医院实施一个强有力的项目对于确保血糖仪的准确性和精密度以产生最佳结果至关重要。本研究的目的是为一家大型三级医疗转诊医院设计一个血糖仪评估的质量保证项目,并评估2018年7月至2019年7月的结果。医院使用的73台血糖仪分别每月进行一次内部质量控制检查,每3个月进行一次能力验证测试。对结果进行审查并绘制在布兰德-奥特曼图上。根据ISO 15197:2013进行克拉克误差网格分析,以评估血糖浓度测量不准确的临床意义。有8台设备被ISO标准判定为不合格,随后进行了更换。在克拉克误差网格分析中,96.83%和3.17%的值分别处于A区和B区。该研究符合要求99%的值落在A区和B区的标准。对该项目一年后的审查及其识别有缺陷血糖仪的能力验证了该模型的有效性。所采用的方法可作为临床环境中血糖仪质量管理的一个范例。

相似文献

6

本文引用的文献

1
Reducing Treatment Errors Through Point-of-Care Glucometer Configuration.通过即时血糖仪配置减少治疗错误。
Jt Comm J Qual Patient Saf. 2018 Nov;44(11):683-694. doi: 10.1016/j.jcjq.2018.03.014. Epub 2018 Jul 9.
8
Glucometers-need for external quality control.血糖仪——外部质量控制的必要性。
Indian J Clin Biochem. 2000 Jul;15(2):128-33. doi: 10.1007/BF02883741.