孟买一家三级护理医院中,医院和门诊环境中使用的血糖仪与实验室参考方法的比较。

Comparison of glucometers used in hospitals and in outpatient settings with the laboratory reference method in a tertiary care hospital in Mumbai.

作者信息

Ullal Aarti, Parmar Girish M, Chauhan Phulrenu H

机构信息

Department of Endocrinology, P.D. Hinduja National Hospital and Medical Research Centre, Veer Savarkar Marg, Mahim, Mumbai, Maharashtra, India.

出版信息

Indian J Endocrinol Metab. 2013 Dec;17(Suppl 3):S688-93. doi: 10.4103/2230-8210.123569.

Abstract

BACKGROUND

Glucometers allow self-monitoring of blood glucose in a convenient manner. With the availability of various glucometers, there is a persistent attempt to improve the accuracy and the precision of these glucometer readings, so as to match the laboratory values of blood glucose.

OBJECTIVE

We compared the glucometers used in hospital and out-patient settings with the laboratory reference method.

MATERIALS AND METHODS

We analyzed a total of 105 blood samples collected from in-patient and out-patient from our tertiary care hospital. Venous blood samples were collected and checked on six glucometers and the same blood sample was sent to the laboratory for glucose estimation. The laboratory value was used as a reference for comparison. The accuracy was evaluated by the ISO criteria. The results were evaluated by Bland Altman graphs, correlation coefficients, scatter plots and Clarke's error grid analysis.

RESULTS

We observed good correlation between bed side glucometer and laboratory automated analyzer. Among the in-patient glucometers Breeze 2, Performa and SureStep, the correlation coefficient was 0.97, 0.96 and 0.88 respectively. Among the outpatient glucometers One touch ultra 2, Active and Contour, the correlation coefficient was 0.97, 0.97 and 0.95 respectively.

CONCLUSIONS

There is a good correlation between different glucometers and laboratory values especially in the out-patient settings. Among all in-patient glucometers, SureStep by Johnson and Johnson had least correlation coefficient, whereas all three out-patient glucometers correlated well with the laboratory values.

摘要

背景

血糖仪能够以便捷的方式实现血糖自我监测。随着各种血糖仪的出现,人们一直在努力提高这些血糖仪读数的准确性和精密度,以使其与血糖的实验室检测值相匹配。

目的

我们将医院和门诊环境中使用的血糖仪与实验室参考方法进行了比较。

材料与方法

我们分析了从我们的三级护理医院的住院患者和门诊患者中采集的总共105份血样。采集静脉血样并在六种血糖仪上进行检测,同时将相同的血样送至实验室进行血糖测定。将实验室检测值用作比较的参考。按照ISO标准评估准确性。通过布兰德-奥特曼图、相关系数、散点图和克拉克误差网格分析对结果进行评估。

结果

我们观察到床边血糖仪与实验室自动分析仪之间具有良好的相关性。在住院患者使用的血糖仪Breeze 2、Performa和SureStep中,相关系数分别为0.97、0.96和0.88。在门诊患者使用的血糖仪One touch ultra 2、Active和Contour中,相关系数分别为0.97、0.97和0.95。

结论

不同血糖仪与实验室检测值之间具有良好的相关性,尤其是在门诊环境中。在所有住院患者使用的血糖仪中,强生公司的SureStep相关系数最低,而所有三种门诊患者使用的血糖仪与实验室检测值的相关性都很好。