由于故障电气装置产生的电磁干扰,导致电击传递不当。

Inappropriate shock delivery as a result of electromagnetic interference originating from the faulty electrical installation.

机构信息

Institute for Cardiovascular Diseases Dedinje, Belgrade, Serbia.

School Of Medicine, University of Belgrade, Belgrade, Serbia.

出版信息

Ann Noninvasive Electrocardiol. 2022 Sep;27(5):e12952. doi: 10.1111/anec.12952. Epub 2022 Apr 25.

Abstract

We present a case report of a 74-year-old male patient with an implantable cardioverter defibrillator who suffered an inappropriate defibrillation shock while bathing in the tub. Insight in the ICD stored electrogram episodes revealed electromagnetic interferences, with a typical 50 Hz electrical artifact mimicking fast ventricular tachycardia as a device misinterpreted. After this event, the maintenance workers investigated the electrical installation in the bathroom and revealed that there was voltage leaking between electrical installation and metal pipes. After the repair was completed without any additional programming, the patient has had no subsequent shocks.

摘要

我们报告了 1 例 74 岁男性患者的病例,该患者在浴缸中洗澡时,其植入式心脏复律除颤器发生了不适当的除颤电击。对 ICD 存储的电描记图事件进行分析,发现存在电磁干扰,典型的 50Hz 电伪影模拟快速室性心动过速,导致设备误判。此次事件发生后,维护人员检查了浴室的电气安装情况,发现电气安装和金属管道之间存在电压泄漏。在完成修复且未进行任何额外程控后,患者未再发生电击。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/760e/9484028/0bdd8a795125/ANEC-27-e12952-g002.jpg

相似文献

1
Inappropriate shock delivery as a result of electromagnetic interference originating from the faulty electrical installation.
Ann Noninvasive Electrocardiol. 2022 Sep;27(5):e12952. doi: 10.1111/anec.12952. Epub 2022 Apr 25.
2
Inappropriate ICD shock due to hot tub-induced external electrical interference.
J Electrocardiol. 2018 Sep-Oct;51(5):852-855. doi: 10.1016/j.jelectrocard.2018.06.016. Epub 2018 Jun 26.
4
Importance of abortive shock capability with electrogram storage in cardioverter-defibrillator devices.
J Am Coll Cardiol. 1993 Mar 15;21(4):895-900. doi: 10.1016/0735-1097(93)90344-z.
7
Multiple inappropriate shocks precipitated by interrogation of an implantable cardioverter defibrillator.
Pacing Clin Electrophysiol. 1998 Jun;21(6):1329-30. doi: 10.1111/j.1540-8159.1998.tb00200.x.
8
Outcomes of defibrillator therapy in catecholaminergic polymorphic ventricular tachycardia.
Heart Rhythm. 2014 Jan;11(1):58-66. doi: 10.1016/j.hrthm.2013.10.027. Epub 2013 Oct 11.
9
Inappropriate implantable cardioverter defibrillator shock from a transcutaneous muscle stimulation device therapy.
J Interv Card Electrophysiol. 2005 Jun;13(1):73-5. doi: 10.1007/s10840-005-0357-3.
10
Development and testing of an algorithm to detect implantable cardioverter-defibrillator lead failure.
Heart Rhythm. 2006 Feb;3(2):155-62. doi: 10.1016/j.hrthm.2005.11.009.

本文引用的文献

2
Inappropriate ICD shock due to hot tub-induced external electrical interference.
J Electrocardiol. 2018 Sep-Oct;51(5):852-855. doi: 10.1016/j.jelectrocard.2018.06.016. Epub 2018 Jun 26.
3
Are patients with cardiac implants protected against electromagnetic interference in daily life and occupational environment?
Eur Heart J. 2015 Jul 21;36(28):1798-804. doi: 10.1093/eurheartj/ehv135. Epub 2015 Apr 22.
4
Troubleshooting implanted cardioverter defibrillator sensing problems I.
Circ Arrhythm Electrophysiol. 2014 Dec;7(6):1237-61. doi: 10.1161/CIRCEP.114.002344.
6
Electromagnetic interference with implantable cardioverter-defibrillators at power frequency: an in vivo study.
Circulation. 2014 Jan 28;129(4):441-50. doi: 10.1161/CIRCULATIONAHA.113.003081. Epub 2013 Oct 25.
10
Guidelines for limiting exposure to time-varying electric and magnetic fields (1 Hz to 100 kHz).
Health Phys. 2010 Dec;99(6):818-36. doi: 10.1097/HP.0b013e3181f06c86.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验