Tecnológico Nacional de México, Instituto Tecnológico de Chihuahua, Ave. Tecnológico No. 2909, Chihuahua Chih. 31310, Mexico.
Facultad de Medicina y Ciencias Biomédicas, Universidad Autónoma de Chihuahua, Circuito Universitario 31109, Chihuahua Chih. 31125, Mexico.
Rev Sci Instrum. 2022 Apr 1;93(4):043101. doi: 10.1063/5.0080572.
In this article, the use of a square Hartmann screen test to measure the radius of curvature of a corneal topography calibration test sphere is presented. The proposed technique is based on the image formation principle by specular reflection on convex reflective surfaces. Applying an inverse Hartmann test, a de-magnified virtual image (Hartmanngram) is obtained; considering their own scaled reference screen plate, a zonal wavefront retrieval approach is used and the radius of curvature obtained. Experimental setup along the obtained results is presented. A simulated spherical wavefront is used as a method to evaluate the error in the wavefront reconstruction. Since the measurements of radius of curvature fits in to ISO 10343, through suitable modifications the proposed method is potentially applicable in small F/# convex specular surfaces, as is the case in keratometry and corneal topography measurements.
本文提出了一种利用方形哈特曼屏测试来测量角膜地形校准测试球的曲率半径的方法。该技术基于镜面反射在凸面反射镜上的成像原理。应用逆哈特曼测试,可以得到一个缩小的虚拟图像(哈特曼网格);考虑到自身的缩放参考屏幕板,可以使用带区波前重建方法来获得曲率半径。本文还介绍了根据所得到的结果进行的实验设置。使用模拟球面波前作为评估波前重建误差的方法。由于曲率半径的测量符合 ISO 10343 标准,因此通过适当的修改,该方法可能适用于小 F/# 凸面镜面,如角膜曲率计和角膜地形测量中。