多动态多回波序列测量 T1、T2 弛豫时间的扫描参数效应:一项体模研究。

The effect of scan parameters on T1, T2 relaxation times measured with multi-dynamic multi-echo sequence: a phantom study.

机构信息

Department of Radiology, Beijing Friendship Hospital, Capital Medical University, Yong An Road 95, Beijing, 100050, China.

Department of Radiology, Beijing ChuiYangLiu Hospital, Beijing, China.

出版信息

Phys Eng Sci Med. 2022 Jun;45(2):657-664. doi: 10.1007/s13246-022-01128-0. Epub 2022 May 13.

Abstract

Multi-Dynamic Multi-Echo (MDME) Sequence is a new method which can acquire various contrast-weighted images using quantitative relaxometric parameters measured from multicontrast images. The purpose of our study was to investigate the effect of scan parameters of MDME Sequence on measured T1, T2 values of phantoms at 3.0 T MRI scanner. Gray matter, white matter and cerebrospinal fluid simulation phantoms with different relaxation times (named GM, WM, CSF, respectively) were used in our study. All the phantoms were scanned 9 times on different days using MDME sequence with variations of echo train length, matrix, and acceleration factor. The T1, T2 measurements were acquired after each acquisition. The repeatability was characterized as the intragroup coefficient of variation (CV) of measured values over 9 times, and the discrepancies of measurements across different groups were characterized as intergroup CVs. The highest intragroup CVs of T1-GM, T2-GM, T1-WM, T2-WM, T1-CSF, T2-SCF were 1.36%, 1.75%, 0.74%, 1.41%, 1.70%, 7.79%, respectively. The highest intergroup CVs of T1-GM, T2-GM, T1-WM, T2-WM, T1-CSF, T2-SCF were 0.54%, 1.86%, 1.70%, 0.94%, 1.00%, 2.17%, respectively. Quantitative T1, T2 measurements of gray matter, white matter and cerebrospinal fluid simulation phantoms derived from the MDME sequence were not obviously affected by variations of scanning parameters, such as echo train length, matrix, and acceleration factor on 3T scanner.

摘要

多动态多回波(MDME)序列是一种新方法,它可以通过从多对比度图像中测量定量弛豫参数来获取各种对比加权图像。我们的研究目的是研究 MDME 序列的扫描参数对 3.0T MRI 扫描仪上的体模测量 T1、T2 值的影响。我们的研究使用了具有不同弛豫时间的灰质、白质和脑脊液模拟体模(分别命名为 GM、WM、CSF)。所有体模均使用 MDME 序列在不同天进行 9 次扫描,变化包括回波链长度、矩阵和加速因子。每次采集后均采集 T1、T2 测量值。通过 9 次测量值的组内变异系数(CV)来描述重复性,通过不同组之间的测量值差异来描述组间 CV。T1-GM、T2-GM、T1-WM、T2-WM、T1-CSF、T2-CSF 的最高组内 CV 分别为 1.36%、1.75%、0.74%、1.41%、1.70%、7.79%。T1-GM、T2-GM、T1-WM、T2-WM、T1-CSF、T2-CSF 的最高组间 CV 分别为 0.54%、1.86%、1.70%、0.94%、1.00%、2.17%。从 MDME 序列得出的灰质、白质和脑脊液模拟体模的定量 T1、T2 测量值不受扫描参数变化(如回波链长度、矩阵和加速因子)的明显影响,这些参数在 3T 扫描仪上。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/b875/9239947/4bafbcd0ea8b/13246_2022_1128_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验