Radiation Protection Unit, University Hospital Zurich, University of Zurich, Zurich, Switzerland.
Qaelum NV, Gaston Geenslaan 9, 3001 Leuven, Belgium.
Phys Med. 2022 Jul;99:10-15. doi: 10.1016/j.ejmp.2022.05.002. Epub 2022 May 19.
Dose management systems (DMS's) have been widely used for radiation dose data collection and analysis, and have been proven useful for optimization. Although DMS's are used for patient dosimetry, a quality assurance (QA) program has not yet been developed. The aim of this work was to propose vendor neutral quality tests and discuss how QA can be established for DMS's to ensure robustness and quality of patient dose data.
MATERIALS & METHODS: Different DMS's functions were organized in three groups according to their complexity. As basic functions, we defined those that depend on simple data collection and require data verification. Intermediate functions provide calculations such as effective dose or higher dose alerts, while the advanced group aims to address complicated analyses, such as peak skin dose calculations.
An Excel file was developed that includes quality tests on basic and intermediate functions for different modalities (radiography, interventional radiology, CT and mammography) and is provided as Annex. For advanced functionalities, tests are discussed to guide medical physics experts to design their QA program as appropriate.
The QA program is intended to ensure the trustworthy functioning for applications in normal circumstances and patient outliers. As DMS's are systems with regular upgrades, the associated QA program should be a dynamic process that takes into account the functionalities of the DMS and its application in the clinics. Medical physics experts should ensure the safety and accuracy of DMS's, as an essential component in patient dosimetry - part of their mission.
剂量管理系统(DMS)已广泛用于辐射剂量数据的收集和分析,并且已被证明有助于优化。尽管 DMS 用于患者剂量测量,但尚未开发出质量保证(QA)程序。本研究旨在提出供应商中立的质量测试,并讨论如何为 DMS 建立 QA,以确保患者剂量数据的稳健性和质量。
根据其复杂性,将不同的 DMS 功能组织成三组。作为基本功能,我们定义了那些依赖于简单数据收集并需要数据验证的功能。中间功能提供了有效剂量或更高剂量警报等计算,而高级组旨在解决复杂的分析,如峰值皮肤剂量计算。
开发了一个 Excel 文件,其中包括针对不同模式(放射照相、介入放射学、CT 和乳房 X 光摄影)的基本和中间功能的质量测试,并作为附件提供。对于高级功能,讨论了测试以指导医学物理专家根据需要设计其 QA 程序。
QA 程序旨在确保在正常情况下和患者异常情况下应用程序的可靠运行。由于 DMS 是具有定期升级的系统,因此相关的 QA 程序应该是一个动态过程,需要考虑 DMS 的功能及其在临床中的应用。医学物理专家应确保 DMS 的安全性和准确性,因为这是患者剂量测量的重要组成部分,也是他们使命的一部分。