Suppr超能文献

一个成分不可能大于整体:多克隆免疫球蛋白对结合/直接胆红素测定的干扰及消除——病例报告

A component could not be larger than the whole, polyclonal immunoglobulin interference in conjugated/direct bilirubin assay and elimination: A case report.

作者信息

Soleimani Neda, Khaleghpanah Saeideh, Mohammadzadeh Sahand

机构信息

Department of Pathology Shiraz Medical School Shiraz University of Medical Sciences Shiraz Iran.

Department of Pathology Shiraz Transplant Center Abu Ali Sina Hospital Shiraz University of Medical Sciences Shiraz Iran.

出版信息

Clin Case Rep. 2022 May 15;10(5):e05870. doi: 10.1002/ccr3.5870. eCollection 2022 May.

Abstract

Analytical interferences (hemolysis, icterus, paraproteinemia, and lipemia) are of great concern in laboratory tests. In our case report, the conjugated bilirubin result of a patient was significantly higher than the total bilirubin. We recommend being more cautious with samples that produce indistinguishable results and attempting to resolve the issues before releasing them.

摘要

分析干扰(溶血、黄疸、副蛋白血症和脂血)在实验室检测中备受关注。在我们的病例报告中,一名患者的结合胆红素结果显著高于总胆红素。我们建议对产生难以区分结果的样本更加谨慎,并在发布结果前尝试解决这些问题。