Abstract

The Bain co-axial circuit is a recent and versatile addition to the semiclosed anaesthetic breathing systems. The relationship between the patient's arterial carbon dioxide tension (PaCO2) and fresh gas flow during intermittent positive pressure ventilation (IPPV) using this circuit has been reassessed. A mean PaCO2 of 33,4 mmHg for 64 patients was recorded using a fresh gas flow of 100 ml/kg/min and a mean PaCO2 of 37,3 mmHg for 55 patients using a fresh gas flow off 70 ml/kg/min.

摘要

贝恩同轴回路是半封闭麻醉呼吸系统中最新且用途广泛的一种。人们重新评估了在使用该回路进行间歇正压通气(IPPV)期间患者动脉二氧化碳分压(PaCO2)与新鲜气流之间的关系。64例患者在新鲜气流为100毫升/千克/分钟时记录的平均PaCO2为33.4毫米汞柱,55例患者在新鲜气流为70毫升/千克/分钟时记录的平均PaCO2为37.3毫米汞柱。