储存在活性储存设备中的人角膜的镜面显微镜检查

Specular Microscopy of Human Corneas Stored in an Active Storage Machine.

作者信息

Garcin Thibaud, Crouzet Emmanuel, Perrache Chantal, Lepine Thierry, Gain Philippe, Thuret Gilles

机构信息

Ophthalmology Department, University Hospital, 42000 Saint-Etienne, France.

Corneal Graft Biology, Engineering and Imaging Laboratory, BiiGC, EA2521, Federative Institute of Research in Sciences and Health Engineering, Faculty of Medicine, Jean Monnet University, Rue de la Marandière, 42000 Saint-Etienne, France.

出版信息

J Clin Med. 2022 May 26;11(11):3000. doi: 10.3390/jcm11113000.

Abstract

PURPOSE

Unlike corneas stored in cold storage (CS) which remain transparent and thin, corneas stored in organoculture (OC) cannot be assessed by specular microscopy (SM), because edema and posterior folds occur during storage and prevent from specular reflection. We previously developed an active storage machine (ASM) which restores the intraocular pressure while renewing the storage medium, thus preventing major stromal edema. Its transparent windows allow multimodal corneal imaging in a closed system.

AIM

to present SM of corneas stored in this ASM.

METHODS

Ancillary study of two preclinical studies on corneas stored for one and three months in the ASM. A prototype non-contact SM was developed (CMOS camera, ×10 objective, collimated LED source, micrometric stage). Five non-overlapping fields (935 × 748 μm) were acquired in exactly the same areas at regular intervals. Image quality was graded according to defined categories (American Cornea Donor Study). The endothelial cell density (ECD) was measured with a center method. Finally, ECD was also compared to Hoechst-stained cell nuclei count (ECD).

RESULTS

The 62 corneas remained thin during storage, allowing SM at all time points without corneal deconditioning. Image quality varied depending on donors and days of control but, overall, in the 1100 images, we observed 55% of excellent and 30% of good quality images. ECD did not differ from ECD ( = 0.084).

CONCLUSIONS

The ASM combines the advantages of CS (closed system) and OC (long-term storage). Specular microscopy is possible at any time in the ASM with a large field of view, making endothelial controls easy and safe.

摘要

目的

与冷藏保存(CS)的角膜保持透明且薄不同,器官培养(OC)保存的角膜无法通过镜面显微镜检查(SM)进行评估,因为在保存过程中会出现水肿和后弹力层褶皱,从而阻碍镜面反射。我们之前开发了一种主动保存机(ASM),它在更新保存介质的同时恢复眼内压,从而防止角膜基质出现严重水肿。其透明窗口允许在封闭系统中进行多模式角膜成像。

目的

展示保存在这种ASM中的角膜的镜面显微镜检查结果。

方法

对两项关于角膜在ASM中保存1个月和3个月的临床前研究的辅助研究。开发了一种原型非接触式镜面显微镜(CMOS相机、×10物镜、准直LED光源、测微计载物台)。以固定间隔在完全相同的区域采集五个不重叠的视野(935×748μm)。根据定义的类别(美国角膜供体研究)对图像质量进行分级。采用中心法测量内皮细胞密度(ECD)。最后,还将ECD与经Hoechst染色的细胞核计数(ECD)进行比较。

结果

62只角膜在保存期间保持薄的状态,在所有时间点都能进行镜面显微镜检查,且角膜未出现状态改变。图像质量因供体和对照天数而异,但总体而言,在1100张图像中,我们观察到55%为优质图像,30%为良好质量图像。ECD与ECD无差异( = 0.084)。

结论

ASM结合了CS(封闭系统)和OC(长期保存)的优点。在ASM中随时都可以进行镜面显微镜检查,视野广阔,使内皮细胞检查变得轻松且安全。