Bassey E J, Dallosso H M, Fentem P H, Irving J M, Patrick J M
Eur J Appl Physiol Occup Physiol. 1987;56(3):323-30. doi: 10.1007/BF00690900.
A small (28 g) mechanical accelerometer has been tested by subjecting it to controlled bench tests consisting of repetitive vertical oscillations on two designs of test rig. The accelerometer's 3-digit display provided a cumulated score with a maximum of 99.9 units. This score was compared with an independent count of the imposed oscillations and found to be linear with time (r = 0.996) and reproducible on retest (coefficient of variation = +/- 1.5%). The sensitivity ranged from 6.2 to 7.4 units/10,000 oscillations. The response was related to the maximal applied acceleration (calculated from the amplitude and frequency of the oscillations on the assumption that they were sinusoidal) and independent of the amplitude and frequency used. The threshold maximal acceleration was less than 2 m s-2 and the response had reached a plateau at 4 m s-2. During field studies the accelerometer was firmly attached over the hip in a waistband where it responded to the vertical accelerations produced by walking. When compared with an independent count of footsteps from a heel-mounted resistance pad the accelerometer score (after calibration) was not significantly different. The mean difference was (0.29 +/- 0.67, S.D.) 10(3) "steps" in a younger group (n = 8, mean age 39 years) and (0.46 +/- 1.08, S.D.) 10(3) "steps" in an older group of women (n = 6, mean age 65 years). Scores of around 10 X 10(3) "steps" can be expected in a day in moderately active young subjects and 40 X 10(3) "steps" in a week in the elderly. Simultaneously recorded scores from both right and left hips wee not significantly different.(ABSTRACT TRUNCATED AT 250 WORDS)
一个小型(28克)机械加速度计已通过在两种试验台设计上进行重复垂直振荡的受控台架试验进行了测试。该加速度计的三位数字显示屏提供了一个累积分数,最高为99.9单位。将该分数与施加振荡的独立计数进行比较,发现其与时间呈线性关系(r = 0.996),并且在重新测试时可重复(变异系数=+/- 1.5%)。灵敏度范围为6.2至7.4单位/10,000次振荡。响应与最大施加加速度有关(根据振荡的幅度和频率假设为正弦波计算得出),并且与所使用的幅度和频率无关。阈值最大加速度小于2米/秒²,响应在4米/秒²时达到平稳状态。在现场研究中,加速度计牢固地固定在腰带中的髋部上方,在那里它对行走产生的垂直加速度做出响应。与安装在脚跟处的电阻垫独立计数的步数相比,加速度计分数(校准后)没有显著差异。在较年轻组(n = 8,平均年龄39岁)中,平均差异为(0.29 +/- 0.67,标准差)×10³“步”,在老年女性组(n = 6,平均年龄65岁)中为(0.46 +/- 1.08,标准差)×10³“步”。在适度活跃的年轻受试者中,一天内预计分数约为10×10³“步”,在老年人中一周内为40×10³“步”。同时记录的左右髋部分数没有显著差异。(摘要截断于250字)