用于测量髋关节活动范围的智能手机测角仪的临床可靠性和可用性。

Clinical reliability and usability of smartphone goniometers for hip range of motion measurement.

作者信息

Takeda Yugo, Furukawa Katsuhiro

机构信息

Iwaki Yumoto Hospital: 6 Daiyama, Joban-Yumoto-cho, Iwaki City, Fukushima 972-8321, Japan.

Faculty of Health Sciences, Iryo Sosei University, Japan.

出版信息

J Phys Ther Sci. 2022 Jun;34(6):433-439. doi: 10.1589/jpts.34.433. Epub 2022 Jun 6.

Abstract

[Purpose] This study aimed to utilize the objective measurements and a survey questionnaire for assessing the intra- and inter-rater reliability, measurement time, and usability of a smartphone application type of goniometer to measure the hip joint angles. [Participants and Methods] Three examiners (physiotherapists) measured the hip joint range of motion using universal and smartphone goniometers on 30 daycare center rehabilitation patients. Reliability was calculated using the intra-class correlation coefficient. The examiners completed a questionnaire survey for assessing the usability of the goniometers. [Results] The intra-rater reliability was high, but the inter-rater reliability was low. Measurement times using the two instruments showed no difference. The usability questionnaire findings suggested that the smartphone goniometer was easier to use than the universal goniometer. [Conclusion] Reliability within the raters was high, but reliability among the raters was low. However, both goniometric devices provided a satisfactory range of motion measurement data when a single evaluator used the same device for all measurements.

摘要

[目的] 本研究旨在利用客观测量和调查问卷来评估一款智能手机应用程序式测角仪测量髋关节角度时的评分者内信度、评分者间信度、测量时间及可用性。[参与者与方法] 三名检查者(物理治疗师)使用通用测角仪和智能手机测角仪,对30名日间护理中心康复患者的髋关节活动范围进行测量。使用组内相关系数计算信度。检查者完成一份关于评估测角仪可用性的问卷调查。[结果] 评分者内信度较高,但评分者间信度较低。使用两种仪器的测量时间无差异。可用性问卷调查结果表明,智能手机测角仪比通用测角仪更易于使用。[结论] 评分者内部的信度较高,但评分者之间的信度较低。然而,当单一评估者使用同一设备进行所有测量时,两种测角设备均能提供令人满意的活动范围测量数据。

相似文献

1
Clinical reliability and usability of smartphone goniometers for hip range of motion measurement.
J Phys Ther Sci. 2022 Jun;34(6):433-439. doi: 10.1589/jpts.34.433. Epub 2022 Jun 6.
5
8
Validity and Reliability of 2 Goniometric Mobile Apps: Device, Application, and Examiner Factors.
J Sport Rehabil. 2016 Dec;25(4):371-379. doi: 10.1123/jsr.2015-0041. Epub 2016 Aug 24.
10
The validity and reliability of DrGoniometer, a smartphone application, for measuring forearm supination.
J Hand Ther. 2019 Jan-Mar;32(1):110-117. doi: 10.1016/j.jht.2018.03.003. Epub 2018 Jul 17.

本文引用的文献

2
Reliability and validity of clinically accessible smartphone applications to measure joint range of motion: A systematic review.
PLoS One. 2019 May 8;14(5):e0215806. doi: 10.1371/journal.pone.0215806. eCollection 2019.
4
The validation of smartphone applications for heart rate measurement.
Ann Med. 2018 Dec;50(8):721-727. doi: 10.1080/07853890.2018.1531144. Epub 2018 Nov 17.
6
Accuracy of Smartphone-Based Pulse Oximetry Compared with Hospital-Grade Pulse Oximetry in Healthy Children.
Telemed J E Health. 2018 Jul;24(7):527-535. doi: 10.1089/tmj.2017.0166. Epub 2017 Dec 7.