Walley V M, Keon W J
J Thorac Cardiovasc Surg. 1987 Jun;93(6):925-33.
The experience of the University of Ottawa Heart Institute with valvular replacement by Ionescu-Shiley bovine pericardial bioprostheses has been recently reviewed. The focus of the present study is to examine the patterns of valve failure in a subgroup of these prostheses, those that required explanation for valve failure caused specifically by cusp wear and tears. A total of 22 aortic and six mitral standard-profile valves failed in this way. Twenty-five of these valves were reassessed and are the subject of this paper. Twenty-eight holes and 56 cusp tears occurred in these valves in locations which, along with coexistent pathologic findings, suggested that the cusp commissural suture/alignment stitch was the point of their origin. The importance of this observation and postulation about the genesis of these cases of valve failure lies in the potential for improvement of this valve's performance by modifications of its manufacture.
渥太华大学心脏研究所使用伊奥内斯库 - 希利牛心包生物假体进行瓣膜置换的经验最近得到了回顾。本研究的重点是检查这些假体中的一个亚组的瓣膜失效模式,即那些需要解释由瓣叶磨损和撕裂导致的瓣膜失效的情况。共有22个主动脉瓣和6个二尖瓣标准轮廓瓣膜以这种方式失效。其中25个瓣膜进行了重新评估,是本文的主题。这些瓣膜中出现了28个孔和56处瓣叶撕裂,其位置连同并存的病理发现表明,瓣叶连合缝线/对齐缝线是其起源点。这一观察结果以及关于这些瓣膜失效病例发生原因的假设的重要性在于,通过改进其制造工艺有可能提高这种瓣膜的性能。