Abstract

BACKGROUND

The pedicle screw system is widely used in spine surgery, and it provides rigid fixation and leads to successful subsequent deformity correction and bony fusion. The standard imaging technique for pedicle screw insertion is two-dimensional images obtained from C-arm-type X-ray fluoroscopy. Artis Zeego is an emerging intraoperative imaging technique that can provide conventional two-dimensional fluoroscopic images and rapid three-dimensional fluoroscopic computed tomography reconstruction imaging. The aim of this study is to compare the insertion accuracies of PS placement using Artis Zeego and conventional 2D X-ray fluoroscopy.

METHODS

In this study, we retrospectively reviewed the postoperative images of thoracolumbar fusion patients who underwent surgery using pedicle screws between 2013 and 2018. Pedicle screw malplacement was assessed using a four-grade classification by Rao et al. Misplacement rates were compared between pedicle screws assisted with Artis Zeego and two-dimensional fluoroscopy.

RESULTS

A total of 1107 pedicle screws in 153 patients were inserted using Artis Zeego, and 427 pedicle screws in 80 patients were inserted using fluoroscopy. The overall perforation rate was 4.2% (46 perforations of 1106 pedicle screws) in the Artis Zeego group and 7.7% (33 perforations of 427 pedicle screws) in the fluoroscopy group. In the Artis Zeego group, 43 (3.9%) screws were classified as grade 1, and three (0.3%) screws were classified as grade 2. In the fluoroscopy group, 21 (4.9%) screws were classified as grade 1, 10 (2.3%) screws were classified as grade 2, and 2 (0.5%) screws were classified as grade 3. The use of Artis Zeego was associated with a significantly lower screw malplacement rate than was the use of fluoroscopy (p < 0.001).

CONCLUSIONS

Our results demonstrated that pedicle screw placement with Artis Zeego was associated with a lower malplacement rate than was conventional two-dimensional fluoroscopy. No severe malplacement was observed in the Artis Zeego group. Thus, Artis Zeego could be a good option for improving pedicle screw accuracy.

摘要

背景

椎弓根螺钉系统广泛应用于脊柱外科,它提供了刚性固定,并导致后续成功的畸形矫正和骨融合。椎弓根螺钉插入的标准成像技术是从 C 臂 X 射线透视机获得的二维图像。Artis Zeego 是一种新兴的术中成像技术,它可以提供常规的二维透视图像和快速的三维透视计算机断层扫描重建成像。本研究的目的是比较使用 Artis Zeego 和传统二维 X 射线透视的 PS 放置的插入精度。

方法

在这项研究中,我们回顾性地审查了 2013 年至 2018 年间接受椎弓根螺钉融合手术的胸腰椎融合患者的术后图像。Rao 等人使用四级分类法评估椎弓根螺钉的错位情况。比较了 Artis Zeego 和二维透视辅助下的椎弓根螺钉的错位率。

结果

共有 153 例患者使用 Artis Zeego 插入 1107 枚椎弓根螺钉,80 例患者使用透视法插入 427 枚椎弓根螺钉。Artis Zeego 组总体穿孔率为 4.2%(1106 枚椎弓根螺钉中有 46 枚穿孔),透视组为 7.7%(427 枚椎弓根螺钉中有 33 枚穿孔)。在 Artis Zeego 组中,43 枚(3.9%)螺钉被归类为 1 级,3 枚(0.3%)螺钉被归类为 2 级。在透视组中,21 枚(4.9%)螺钉被归类为 1 级,10 枚(2.3%)螺钉被归类为 2 级,2 枚(0.5%)螺钉被归类为 3 级。与透视法相比,使用 Artis Zeego 螺钉的螺位错率显著降低(p < 0.001)。

结论

我们的结果表明,与传统二维透视相比,使用 Artis Zeego 进行椎弓根螺钉放置的螺位错率较低。在 Artis Zeego 组中没有观察到严重的螺位错。因此,Artis Zeego 可能是提高椎弓根螺钉准确性的一个很好的选择。