Djurovic Koprivica Daniela, Puskar Tatjana, Budak Igor, Sokac Mario, Jeremic Knezevic Milica, Maletin Aleksandra, Milekic Bojana, Vukelic Djordje
Faculty of Medicine, University of Novi Sad, 21000 Novi Sad, Serbia.
Faculty of Technical Sciences, University of Novi Sad, 21000 Novi Sad, Serbia.
Polymers (Basel). 2022 Jul 11;14(14):2821. doi: 10.3390/polym14142821.
The paper presents the influence of impression methods, polymer materials, and implant angulation on the accuracy of the definitive working model for the production of implant-supported dental restorations, based on the analysis of results obtained using different impression methods, materials, and parallel and angulated implants. The study findings indicate that all aforementioned factors impact the accuracy of the definitive working model. Specifically, 20° implant angulation in relation to the vertical plane has a greater impact on the impression accuracy compared to parallel implants. The open and splint method in combination with addition silicone, as well as the splint method and polyether combination yielded more accurate results when using implants under 20° angulation compared to other method and material combinations. The splint method in combination with addition silicone resulted in the smallest mean deviations from the center of the parallel implant base compared to other combinations of methods and materials. Analysis results further revealed statistically significant differences in the measured indicators across impression methods, implants, and polymer materials.
本文基于对使用不同印模方法、材料以及平行和有角度种植体所获得结果的分析,阐述了印模方法、高分子材料和种植体角度对种植体支持式牙修复体制作中最终工作模型准确性的影响。研究结果表明,上述所有因素均会影响最终工作模型的准确性。具体而言,与平行种植体相比,种植体相对于垂直平面呈20°角时对印模准确性的影响更大。与其他方法和材料组合相比,开放和夹板法与加成型硅橡胶联合使用,以及夹板法与聚醚联合使用,在种植体角度小于20°时能产生更准确的结果。与其他方法和材料组合相比,夹板法与加成型硅橡胶联合使用时,与平行种植体基台中心的平均偏差最小。分析结果进一步显示,在印模方法、种植体和高分子材料的各项测量指标上存在统计学显著差异。