铅-212的定量单光子发射计算机断层扫描/计算机断层扫描成像:一项体模研究。

Quantitative SPECT/CT imaging of lead-212: a phantom study.

作者信息

Kvassheim Monika, Revheim Mona-Elisabeth R, Stokke Caroline

机构信息

Department of Physics and Computational Radiology, Division of Radiology and Nuclear Medicine, Oslo University Hospital, Oslo, Norway.

Faculty of Medicine, University of Oslo, Oslo, Norway.

出版信息

EJNMMI Phys. 2022 Aug 4;9(1):52. doi: 10.1186/s40658-022-00481-z.

Abstract

BACKGROUND

Lead-212 (Pb) is a promising radionuclide for targeted therapy, as it decays to α-particle emitter bismuth-212 (Bi) via β-particle emission. This extends the problematic short half-life of Bi. In preparation for upcoming clinical trials with Pb, the feasibility of quantitative single photon-emission computed tomography/computed tomography (SPECT/CT) imaging of Pb was studied, with the purpose to explore the possibility of individualised patient dosimetric estimation.

RESULTS

Both acquisition parameters (combining two different energy windows and two different collimators) and iterative reconstruction parameters (varying the iterations x subsets between 10 × 1, 15 × 1, 30 × 1, 30 × 2, 30 × 3, 30 × 4, and 30 × 30) were investigated to evaluate visual quality and quantitative uncertainties based on phantom images. Calibration factors were determined using a homogeneous phantom and were stable when the total activity imaged exceeded 1 MBq for all the imaging protocols studied, but they increased sharply as the activity decayed below 1 MBq. Both a 20% window centred on 239 keV and a 40% window on 79 keV, with dual scatter windows of 5% and 20%, respectively, could be used. Visual quality at the lowest activity concentrations was improved with the High Energy collimator and the 79 keV energy window. Fractional uncertainty in the activity quantitation, including uncertainties from calibration factors and small volume effects, in spheres of 2.6 ml in the NEMA phantom was 16-21% for all protocols with the 30 × 4 filtered reconstruction except the High Energy collimator with the 239 keV energy window. Quantitative analysis was possible both with and without filters, but the visual quality of the images improved with a filter.

CONCLUSIONS

Only minor differences were observed between the imaging protocols which were all determined suitable for quantitative imaging of Pb. As uncertainties generally decreased with increasing iterative updates in the reconstruction and recovery curves did not converge with few iterations, a high number of reconstruction updates are recommended for quantitative imaging.

摘要

背景

铅 - 212(Pb)是一种有前景的用于靶向治疗的放射性核素,因为它通过β粒子发射衰变为α粒子发射体铋 - 212(Bi)。这延长了Bi存在问题的短半衰期。在为即将开展的Pb临床试验做准备时,研究了Pb定量单光子发射计算机断层扫描/计算机断层扫描(SPECT/CT)成像的可行性,目的是探索个体化患者剂量测定估计的可能性。

结果

研究了采集参数(结合两种不同的能量窗和两种不同的准直器)和迭代重建参数(将迭代次数×子集数在10×1、15×1、30×1、30×2、30×3、30×4和30×30之间变化),以基于体模图像评估视觉质量和定量不确定性。使用均匀体模确定校准因子,对于所研究的所有成像方案,当成像的总活度超过1 MBq时校准因子是稳定的,但当活度衰减到低于1 MBq时它们会急剧增加。可以使用以239 keV为中心的20%窗和79 keV的40%窗,分别具有5%和20%的双散射窗。使用高能准直器和79 keV能量窗时,最低活度浓度下的视觉质量得到改善。对于NEMA体模中2.6 ml球体,除了使用239 keV能量窗的高能准直器外,所有采用30×4滤波重建的方案,活度定量的分数不确定性(包括来自校准因子和小体积效应的不确定性)为16% - 21%。有无滤波都可以进行定量分析,但使用滤波器时图像的视觉质量得到改善。

结论

在所有被确定适用于Pb定量成像的成像方案之间仅观察到微小差异。由于不确定性通常随着重建中迭代更新次数的增加而降低,并且恢复曲线在迭代次数较少时不会收敛,因此建议进行大量的重建更新以进行定量成像。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/19c6/9352840/763d4c02f0e4/40658_2022_481_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索