Canaud B, Nguyen Q V, Argiles A, Polito C, Polascheoo H D, Mion C
Artif Organs. 1987 Apr;11(2):188-90. doi: 10.1111/j.1525-1594.1987.tb02655.x.
To reduce the cost of hemodiafiltration (HDF), a standard hemodialysis machine was modified to permit in-line production of substitution fluid from bicarbonate dialysate. The authors' present data showing that this procedure was made bacteriologically safe; was simple to use, and offered a cheap approach to perform high-flux HDF using a hemodifilter of relatively small membrane surface area.