Bush V, Reed R G
Clin Chem. 1987 Jun;33(6):821-3.
We used automated and manual versions of two commonly used dye-binding methods for albumin to quantify albumin that had been modified by covalent addition of bilirubin. Dye-binding methods based on reaction with bromcresol purple underestimated the modified albumin by as much as 29%. Dye-binding methods based on reaction with bromcresol green were unaffected by the presence of covalently attached bilirubin. Neither method was affected by the presence of noncovalently bound bilirubin.
我们使用了两种常用的白蛋白染料结合法的自动化和手动版本,以量化通过胆红素的共价加成而被修饰的白蛋白。基于与溴甲酚紫反应的染料结合法低估了修饰后的白蛋白,低估幅度高达29%。基于与溴甲酚绿反应的染料结合法不受共价连接胆红素存在的影响。两种方法均不受非共价结合胆红素存在的影响。