Suppr超能文献

用溴甲酚紫测定血清白蛋白的新型自动染料结合法。

New automated dye-binding method for serum albumin determination with bromcresol purple.

作者信息

Pinnell A E, Northam B E

出版信息

Clin Chem. 1978 Jan;24(1):80-6.

PMID:618669
Abstract

We describe a new automated dye-binding method for serum albumin determination with bromcresol purple (BCP) that has several advantages over an existing bromcresol green (BCG) method. The continuous-flow method is sensitive, linear, and precise, with negligible sample interaction at an analytical rate of 60 samples per hour. Unlike BCG, BCP did not react with an albumin-free serum globulin preparation or pure human transferrin solutions. Reaction with serum was instantaneous; in contrast, BCG exhibits a slow nonspecific reaction with some specimens. The specificity of BCP was demonstrated by good agreement with results of "rocket" immunoelectrophoresis (EIA) where y(BCP) = 0.95X(EIA) + 1.72. The BCG method overestimated serum albumin concentration where y(BCG) = 1.01X(EIA) + 6.77. Precipitation, which affects the BCG method, was not observed with BCP. Blank corrections were negligible, salicylate did not interfere, and bilirubin affected the method only if present in very high concentration. The method offers a solution to the poor accuracy of existing BCG methods while retaining many of their desirable features.

摘要

我们描述了一种用溴甲酚紫(BCP)测定血清白蛋白的新型自动化染料结合法,该方法相对于现有的溴甲酚绿(BCG)法具有多个优点。连续流动法灵敏、线性且精确,在每小时分析60个样本的速率下,样本间相互作用可忽略不计。与BCG不同,BCP不与无白蛋白的血清球蛋白制剂或纯人转铁蛋白溶液发生反应。与血清的反应是瞬时的;相比之下,BCG与一些样本会发生缓慢的非特异性反应。BCP的特异性通过与“火箭”免疫电泳(EIA)结果的良好一致性得到证明,其中y(BCP) = 0.95X(EIA) + 1.72。BCG法高估了血清白蛋白浓度,其中y(BCG) = 1.01X(EIA) + 6.77。BCP未观察到影响BCG法的沉淀现象。空白校正可忽略不计,水杨酸盐不产生干扰,胆红素只有在浓度非常高时才会影响该方法。该方法在保留现有BCG法许多理想特性的同时,解决了其准确性差的问题。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验