Department of Radiology, University of Michigan Medical School, Ann Arbor, Michigan, USA.
Department of Radiology, University of Michigan Medical School, Ann Arbor, Michigan, USA.
Ultrasound Med Biol. 2022 Nov;48(11):2276-2291. doi: 10.1016/j.ultrasmedbio.2022.06.020. Epub 2022 Aug 25.
This study investigated induction of pulmonary capillary hemorrhage (PCH) in neonatal pigs (piglets) using three different machines: a GE Venue R1 point-of-care system with C1-5 and L4-12t probes, a GE Vivid 7 Dimension with a 7L probe and a SuperSonic Imagine machine with an SL15-4 probe and shear wave elastography (SWE). Female piglets were anesthetized, and each was mounted vertically in a warm bath for scanning at two or three intercostal spaces. After aiming at an innocuous output, the power was raised for a test exposure. Hydrophone measurements were used to calculate in situ values of mechanical index (MI). Inflated lungs were removed and scored for PCH area. For the C1-5 probe at 50% and 100% acoustical output (AO), a PCH threshold of 0.53 MI was obtained by linear regression (r = 0.42). The L4-12t probe did not induce PCH, but the 7L probe induced zones of PCH in the scan planes. The Venue R1 automated B-line tool applied with the C1-5 probe did not detect PCH induced by the C1-5 probe as B-line counts. However, when PCH induced by C1-5 and 7L exposures were subsequently scanned with the L4-12t probe using the automated tool, B-lines were counted in association with the PCH. The SWE induced PCH at push-pulse positions for 3, 30 and 300 s of SWE with PCH accumulating at 0.33 mm/s and an exponential rise to a maximum of 18.4 mm (r = 0.61). This study demonstrated the induction of PCH by LUS of piglets, and supports the safety recommendation for use of MIs <0.4 in neonatal LUS.
本研究使用三种不同的机器(GE Venue R1 床边系统配 C1-5 和 L4-12t 探头、GE Vivid 7 维度系统配 7L 探头以及 SuperSonic Imagine 机器配 SL15-4 探头和剪切波弹性成像[SWE])在新生仔猪(小猪)中诱导肺毛细血管出血(PCH)。将雌性小猪麻醉,然后将其每只垂直安装在温水浴中,在两个或三个肋间进行扫描。在瞄准无害输出后,提高功率进行测试曝光。使用水听器测量来计算原位机械指数(MI)值。取出充气的肺并对 PCH 面积进行评分。对于 C1-5 探头在 50%和 100%声输出(AO),通过线性回归(r=0.42)获得 0.53 MI 的 PCH 阈值。L4-12t 探头未诱导 PCH,但 7L 探头在扫描平面上诱导 PCH 区。Venue R1 自动 B 线工具与 C1-5 探头一起应用时,未检测到 C1-5 探头诱导的 PCH 作为 B 线计数。然而,当使用自动工具用 L4-12t 探头扫描 C1-5 和 7L 曝光诱导的 PCH 时,与 PCH 相关地计数了 B 线。SWE 在 SWE 的 3、30 和 300 秒推脉冲位置诱导 PCH,PCH 以 0.33 mm/s 的速度积累,并呈指数上升至最大值 18.4 mm(r=0.61)。本研究证明了通过 LUS 诱导小猪 PCH,并支持在新生儿 LUS 中使用 MI<0.4 的安全性建议。