Roccabruna Jacob R, Bridger Karina G, Baran Timothy M
Department of Biomedical Engineering, University of Rochester, Rochester, NY, United States of America.
Department of Imaging Sciences, University of Rochester Medical Center, Rochester, NY, United States of America.
J Mod Opt. 2022;69(12):699-704. doi: 10.1080/09500340.2022.2074159. Epub 2022 May 11.
Quantitative fluorescence spectroscopy requires corresponding reflectance measurements to correct for tissue absorption and scattering. However, it is unclear whether fluorescence adds value beyond the diffuse reflectance measurements necessary for correction. The goal of this study was to compare the accuracy of fluorescence and diffuse reflectance spectroscopy in recovering the concentration of a high-extinction fluorophore, methylene blue (MB), using a compact fiber-optic probe. Fluorescence and diffuse reflectance measurements of tissue simulating phantoms were made using a fiber-optic probe with source-detector separations of 288-1300 μm. Average error in recovered fluorophore concentration was 20.4% for fluorescence and 15.0% for reflectance, though this difference was not significant (p=0.77). Both methods returned concentrations that were similar to known MB concentrations (p≥0.79 in both cases). Fluorescence quantification of the concentration of a high extinction fluorophore did not significantly improve accuracy relative to diffuse reflectance. Investigators should consider whether fluorescence measurements are necessary for a given application.
定量荧光光谱法需要进行相应的反射率测量,以校正组织吸收和散射。然而,尚不清楚荧光是否能在校正所需的漫反射测量之外增加价值。本研究的目的是使用紧凑型光纤探头,比较荧光光谱法和漫反射光谱法在恢复高消光荧光团亚甲蓝(MB)浓度方面的准确性。使用源探测器间距为288 - 1300μm的光纤探头对组织模拟体模进行荧光和漫反射测量。荧光法恢复荧光团浓度的平均误差为20.4%,反射率法为15.0%,不过这种差异并不显著(p = 0.77)。两种方法得到的浓度均与已知的MB浓度相似(两种情况下p≥0.79)。相对于漫反射,高消光荧光团浓度的荧光定量并未显著提高准确性。研究人员应考虑荧光测量对于特定应用是否必要。