Radiological Physics and Advisory Division, Bhabha Atomic Research Centre, Mumbai 400085, India.
Biomed Phys Eng Express. 2022 Sep 13;8(6). doi: 10.1088/2057-1976/ac8e71.
The purpose of this study was to evaluate the performance of solid-state sensor based computed tomography dose profiler (CTDP) probe for measurement of standard computed tomography dose metric CTDIand free in air geometric efficiency for various beam widths available in a 128-slice CT scanner and also to estimate the efficiency of CTDImetric. The response accuracy of CTDP probe was verified using a standard 100 mm long ionization chamber. The geometric efficiency measurements performed by the CTDP probe were validated using XR-QA2 radiochromic film measurements. The efficiency of CTDImetric was assessed by calculating the ratio of CTDImeasured in the center hole position to CTDImeasured in the same position of both head and body phantoms. The weighted CTDI values derived from CTDImeasured by the CTDP probe showed an average difference of 8% from ionization chamber measured values. The efficiency of CTDImetric estimated using CTDP probe and 150 mm long phantoms was in the range of 82% to 86% and 76% to 80% for head and body phantom measurements respectively. The differences in the geometric efficiency values for various beam settings and tube voltages measured by the CTDP probe and films were within 7%. Taken together, the results of this study proved that unlike the 100 mm long ionization chamber, the CTDP probe can be efficiently used to determine CTDI for any length over which dose integration is desired and also measure geometric efficiency of MDCT scanners for various beam widths in helical mode of operation.
本研究旨在评估基于固态传感器的计算机断层扫描剂量分布仪(CTDP)探头在测量标准 CT 剂量学参数 CTDI 方面的性能,以及在 128 层 CT 扫描仪中各种射束宽度下自由空气几何效率,并估计 CTDI 测量的效率。使用标准的 100mm 长电离室验证了 CTDP 探头的响应准确性。使用 XR-QA2 光致变色胶片测量验证了 CTDP 探头进行的几何效率测量。通过计算在中心孔位置测量的 CTDI 与在头部和身体体模相同位置测量的 CTDI 之比,评估了 CTDI 测量的效率。从 CTDP 探头测量的 CTDI 得出的加权 CTDI 值与电离室测量值平均相差 8%。使用 CTDP 探头和 150mm 长体模估算的 CTDI 测量效率分别在头部和身体体模测量中为 82%至 86%和 76%至 80%。CTDP 探头和胶片测量的各种射束设置和管电压的几何效率值之间的差异在 7%以内。总的来说,这项研究的结果证明,与 100mm 长的电离室不同,CTDP 探头可以有效地用于确定所需剂量积分的任何长度的 CTDI,并且还可以测量各种射束宽度下螺旋模式运行的 MDCT 扫描仪的几何效率。