Interobserver Agreement of Electrode to Retina Distance Measurements in a Second-Generation (44-Channel) Suprachoroidal Retinal Prosthesis.

机构信息

Centre for Eye Research Australia, Royal Victorian Eye and Ear Hospital, Victoria, Australia.

Department of Surgery (Ophthalmology), University of Melbourne, Victoria, Australia.

出版信息

Transl Vis Sci Technol. 2022 Sep 1;11(9):4. doi: 10.1167/tvst.11.9.4.

Abstract

PURPOSE

The electrode to retina (ER) distance is an important contributory factor to the safety and efficacy of a suprachoroidal retinal prosthesis. Measuring ER distance may be performed by different observers during multisite studies. The aim of this study was to assess the interobserver agreement in measuring ER distance.

METHODS

Three independent, trained observers measured ER distance from the center of each suprachoroidal electrode to the inner retinal pigment epithelium in spectral-domain optical coherence tomography (SD-OCT) B-scans. A total of 121 ER distance measurements from 77 B-scans collected over 5 months from one subject implanted with a second-generation 44-channel suprachoroidal retinal prosthesis (NCT03406416) were made by each observer.

RESULTS

ER distance ranged from 208 to 509 µm. Pearson's correlation coefficient (ρ) showed agreement of 0.99 (95% confidence interval [CI] = 0.98-0.99) in measuring ER for each pairwise comparison. The mean difference in ER distance between observers ranged from 2.4 to 6.4 µm with pairwise limits of agreement (95% CI) of ±20 µm (5.5% of mean). Intraclass correlation coefficient (ICC) showed agreement of 0.98 (95% CI = 0.97-0.99) between observers.

CONCLUSIONS

There is high agreement in measuring ER distances for suprachoroidal retinal prostheses using our systematic approach between multiple, trained observers, supporting the use of a single observer for each image.

TRANSLATIONAL RELEVANCE

High interobserver agreement outcomes indicate that multiple, trained observers can be used to take ER measurements across different images in suprachoroidal retinal prosthesis studies. This improves multisite study efficiency and gives confidence in interpreting results relating to the safety and efficacy of suprachoroidal retinal prostheses.

摘要

目的

电极到视网膜(ER)的距离是影响脉络膜上视网膜假体安全性和有效性的一个重要因素。在多中心研究中,可能会由不同的观察者来测量 ER 距离。本研究旨在评估测量 ER 距离的观察者间一致性。

方法

3 名独立的、受过训练的观察者在频域光学相干断层扫描(SD-OCT)B 扫描中,从每个脉络膜上电极的中心测量到内视网膜色素上皮的 ER 距离。每位观察者共对 1 名植入第二代 44 通道脉络膜上视网膜假体(NCT03406416)的受试者在 5 个月期间收集的 77 个 B 扫描中的 121 个 ER 距离进行了测量。

结果

ER 距离范围为 208 至 509μm。Pearson 相关系数(ρ)显示,对于每个两两比较,测量 ER 的一致性为 0.99(95%置信区间[CI] = 0.98-0.99)。观察者之间 ER 距离的平均差异范围为 2.4 至 6.4μm,双边一致性界限(95%CI)为±20μm(平均的 5.5%)。组内相关系数(ICC)显示观察者之间的一致性为 0.98(95%CI = 0.97-0.99)。

结论

使用我们的系统方法,多名受过训练的观察者之间在测量脉络膜上视网膜假体的 ER 距离方面具有高度一致性,支持为每个图像使用单个观察者。