Palmer C
Brain Res Bull. 1978 May-Jun;3(3):285-9. doi: 10.1016/0361-9230(78)90129-6.
A technique is described in which bundles of 25 mu dia. insulated wires can be stereotaxically implanted for single unit recording in the brain. With this method extremely stable recordings can be obtained lasting many hours and sometimes days. The unit activity recorded with these electrodes has been found comparable with that recorded from rigid etched microelectrodes in conscious animals. The fine wire electrodes have also been found to be durable for at least a month when chronically implanted in cats. The electrode assembly is easy to construct and the electrodes are easily implanted in any region of the brain.
本文描述了一种技术,通过该技术可将直径为25微米的绝缘线束进行立体定位植入,用于大脑中的单神经元记录。使用这种方法能够获得极其稳定的记录,持续数小时甚至数天。已发现用这些电极记录的神经元活动与在清醒动物中使用刚性蚀刻微电极记录的活动相当。还发现,当长期植入猫体内时,细线电极至少可持续使用一个月。电极组件易于构建,电极也易于植入大脑的任何区域。