Measurement of ventilatory system compliance in infants and young children.

作者信息

Marchal F, Crance J P

出版信息

Respir Physiol. 1987 Jun;68(3):311-8. doi: 10.1016/s0034-5687(87)80016-6.

Abstract

Ventilatory system compliance (Crs) was measured in 32 sedated sleeping infants and young children aged 3-54 months breathing spontaneously through a face mask. Airways were occluded during expiration at different lung volumes above FRC and the corresponding airway pressure was measured. A pressure-volume relationship was then calculated by the least square method. To ensure the presence of the Breuer-Hering inspiratory inhibitory reflex, the percent increase of occluded expiratory time relative to the preceding unoccluded breath was calculated in 28 subjects where several end inspiratory occlusions had been maintained up to the next inspiratory effort. This index ranged from 4 to 100% and correlated negatively with age (r = -0.50, P less than 0.01). Crs ranged from 4.5 to 21.8 ml/cm H2O and correlated best with height (H) (Crs = 5.36 X 10(-4) H2.27, r = 0.94).

摘要

在32名3至54个月大的镇静睡眠婴幼儿中,通过面罩自主呼吸时测量通气系统顺应性(Crs)。在高于功能残气量(FRC)的不同肺容量下呼气时气道被阻塞,并测量相应的气道压力。然后用最小二乘法计算压力-容量关系。为确保存在布雷尔-黑林吸气抑制反射,在28名受试者中计算了阻塞呼气时间相对于前一次未阻塞呼吸的百分比增加,这些受试者中有几次吸气末阻塞一直持续到下一次吸气努力。该指数范围为4%至100%,与年龄呈负相关(r = -0.50,P < 0.01)。Crs范围为4.5至21.8 ml/cm H₂O,与身高(H)的相关性最佳(Crs = 5.36×10⁻⁴H².²⁷,r = 0.94)。