College of Optometry, The Ohio State University, Columbus, Ohio, USA.
College of Optometry, University of Houston, Houston, Texas, USA.
Ophthalmic Physiol Opt. 2023 Jan;43(1):64-72. doi: 10.1111/opo.13054. Epub 2022 Sep 26.
The purpose of this study was to determine intrasession repeatability of a worksheet style contrast sensitivity test (SpotChecks) in children and agreement with an established contrast sensitivity test (Pelli-Robson).
Forty-three children aged 4 to 12 years participated in this single visit study that included two administrations of the SpotChecks binocularly, a single administration of the Pelli-Robson test and other measures of visual performance such as high-contrast visual acuity. Test order was randomised, and participants wore their habitual correction (39 unaided, 4 wearing glasses) for testing. Bland-Altman plots were used to assess the test-retest repeatability of SpotChecks and its agreement with the Pelli-Robson test. Multiple linear regressions were performed to evaluate whether contrast sensitivity was related to participant characteristics such as age, sex and near binocular visual acuity.
The mean difference in log contrast sensitivity (logCS) between two administrations of the SpotChecks was 0.01, with a coefficient of repeatability (1.96*SD of differences) of 0.14 logCS. The mean difference between SpotChecks and Pelli-Robson was 0.00 logCS with 95% limits of agreement of -0.19 to +0.20. For both tests, a statistically significant increase in logCS was associated with age (slopes were 0.02 logCS/year, p < 0.001 and 0.01 logCS/year, p = 0.02 for the SpotChecks and Pelli-Robson tests, respectively).
The SpotChecks test shows good intrasession repeatability and excellent agreement with the Pelli-Robson test in children. Contrast sensitivity showed an increase in logCS with age in children for both tests.
本研究旨在确定工作表式对比敏感度测试(SpotChecks)在儿童中的日内重复性,以及与已建立的对比敏感度测试(Pelli-Robson)的一致性。
43 名 4 至 12 岁的儿童参与了这项单访视研究,包括两次双眼接受 SpotChecks 测试、一次单眼接受 Pelli-Robson 测试以及其他视觉表现测量,如高对比度视力。测试顺序随机,参与者在测试时佩戴其常规矫正(39 名不戴眼镜,4 名戴眼镜)。Bland-Altman 图用于评估 SpotChecks 的测试-再测试重复性及其与 Pelli-Robson 测试的一致性。进行多元线性回归分析,以评估对比敏感度是否与参与者特征(如年龄、性别和近用双眼视力)相关。
两次 SpotChecks 测试的 log 对比敏感度(logCS)平均值差异为 0.01,重复性系数(差异的 1.96*SD)为 0.14 logCS。SpotChecks 与 Pelli-Robson 之间的平均差异为 0.00 logCS,95%一致性界限为-0.19 至 +0.20。对于这两种测试,logCS 均与年龄呈显著正相关(斜率分别为 0.02 logCS/年,p<0.001 和 0.01 logCS/年,p=0.02,分别用于 SpotChecks 和 Pelli-Robson 测试)。
SpotChecks 测试在儿童中具有良好的日内重复性和与 Pelli-Robson 测试的极好一致性。两种测试的 logCS 均随年龄增长而增加。