Department of Paediatric Ophthalmology and Adult Strabismus Services, Aravind Eye Hospital, Madurai, Tamil Nadu, India.
Indian J Ophthalmol. 2022 Oct;70(10):3625-3628. doi: 10.4103/ijo.IJO_487_22.
The main objective is to test the measurements made by an automated eye-tracking system in the presence of strabismus and to compare the data with manual measurements of deviation.
A prospective observational cross-sectional masked double-blinded study was conducted in a tertiary eye care center with 39 participants included in our study, aged 3-41 years. Initial screening of all participants was performed by an ophthalmologist. Ocular deviations were evaluated and compared between manual measurements and an automated eye-tracking system. The device is based on eye-tracking technology. The participants had either a congenital or acquired type of manifest or latent strabismus. Children less than 3 years of age, visual acuity <6/36, and abnormal configuration of the anterior segment were excluded from the study.
The prism alternate cover test (PACT) manual measurements and the automated alternate cover test for measuring horizontal deviation, the manual measurement, and the automated eye track system showed a highly positive correlation (r = 0.932, P < 0.001). The Bland Altman plot analysis shows good agreement between the two measurements, with the mean difference between the two measurements being 1.55 PD, and the 95% limit of agreement was ± 10 PD.
The results obtained with an automated eye-tracking system correlate well with manual strabismus measurements with PACT in terms of diagnosis, precision, and accuracy, with an added benefit of lesser time consumption in performing the test in cooperative/motivated patients. Considering these aspects, patients above the age of 3 years could be assessed with the equipment.
本研究的主要目的是测试斜视患者的自动眼动跟踪系统的测量结果,并将其与偏斜手动测量数据进行比较。
在一家三级眼科中心进行了一项前瞻性观察性横断面、双盲、对照研究,共纳入 39 名年龄在 3 至 41 岁的研究对象。所有参与者均由眼科医生进行初步筛查。评估并比较了手动测量和自动眼动跟踪系统之间的眼位偏差。该设备基于眼动跟踪技术。参与者患有先天性或获得性显性或隐性斜视。3 岁以下儿童、视力<6/36 和眼前节异常者被排除在研究之外。
棱镜交替遮盖试验(PACT)手动测量值和水平偏差的自动交替遮盖试验、手动测量值和自动眼跟踪系统之间呈高度正相关(r=0.932,P<0.001)。Bland-Altman 图分析显示两种测量方法之间具有良好的一致性,两种测量方法之间的平均差异为 1.55 PD,95%一致性界限为±10 PD。
自动眼动跟踪系统的测量结果与手动斜视测量(使用 PACT)在诊断、精密度和准确性方面具有良好的相关性,并且在对合作/有动机的患者进行测试时,该设备具有时间消耗更少的优势。考虑到这些方面,3 岁以上的患者可以使用该设备进行评估。