Suppr超能文献

内斜视儿童棱镜交替遮盖试验的观察者间可靠性

Interobserver reliability of the prism and alternate cover test in children with esotropia.

出版信息

Arch Ophthalmol. 2009 Jan;127(1):59-65. doi: 10.1001/archophthalmol.2008.548.

Abstract

OBJECTIVE

To determine 95% limits of agreement on a measurement and on a difference between 2 measurements for the prism and alternate cover test (PACT) at distance and at near fixation.

METHODS

In a test-retest reliability study, 143 children aged 60 months or younger with esotropia were examined by 2 masked examiners on 1 or 2 occasions yielding 199 test-retest pairs for PACT at distance fixation and 239 test-retest pairs for PACT at near fixation.

RESULTS

For angles greater than 20 prism diopters (PD), the 95% limits of agreement on a measurement and on a difference between 2 measurements were +/-7.3 PD and +/-10.4 PD, respectively, for PACT at distance and +/-8.3 PD and +/-11.7 PD, respectively, for PACT at near. For angles of 10 to 20 PD, the 95% limits of agreement on a measurement and on a difference between 2 measurements were +/-4.1 PD and +/-5.8 PD, respectively, for PACT at distance and +/-3.3 PD and +/-4.7 PD, respectively, for PACT at near.

CONCLUSION

In childhood esotropia, differences of 12 PD or more for angles greater than 20 PD and differences of 6 PD or more for angles between 10 PD and 20 PD are likely to indicate real change. Smaller differences could be real change but could also be due to measurement error.

摘要

目的

确定在远距离和近距离注视时,棱镜交替遮盖试验(PACT)测量值及两次测量差值的95%一致性界限。

方法

在一项重测信度研究中,两名掩蔽检查者对143名60个月及以下患有内斜视的儿童进行了1次或2次检查,获得了199对远距离注视时PACT的重测数据对以及239对近距离注视时PACT的重测数据对。

结果

对于大于20棱镜度(PD)的角度,远距离PACT测量值及两次测量差值的95%一致性界限分别为±7.3 PD和±10.4 PD,近距离PACT分别为±8.3 PD和±11.7 PD。对于10至20 PD的角度,远距离PACT测量值及两次测量差值的95%一致性界限分别为±4.1 PD和±5.8 PD,近距离PACT分别为±3.3 PD和±4.7 PD。

结论

在儿童内斜视中,大于20 PD的角度差值达到12 PD或更大,以及10 PD至20 PD之间的角度差值达到6 PD或更大,可能表明存在实际变化。较小的差值可能是实际变化,但也可能是由于测量误差所致。

相似文献

1
Interobserver reliability of the prism and alternate cover test in children with esotropia.
Arch Ophthalmol. 2009 Jan;127(1):59-65. doi: 10.1001/archophthalmol.2008.548.
3
Multiple Factors Causing Variability of Alignment in Childhood Concomitant Strabismus.
Am J Ophthalmol. 2025 Feb;270:77-82. doi: 10.1016/j.ajo.2024.10.013. Epub 2024 Oct 23.
4
Measurement of strabismic angle using the distance Krimsky test.
Korean J Ophthalmol. 2013 Aug;27(4):276-81. doi: 10.3341/kjo.2013.27.4.276. Epub 2013 Jul 18.
5
Prism and Maddox rod test as a reliable tool for surgical targeting in small-angle AACE.
Sci Rep. 2025 Mar 14;15(1):8872. doi: 10.1038/s41598-025-93791-y.
6
Automated analysis of binocular alignment using an infrared camera and selective wavelength filter.
Invest Ophthalmol Vis Sci. 2013 Apr 17;54(4):2733-7. doi: 10.1167/iovs.12-11400.
7
Efficacy of anti-suppression therapy in improving binocular vision in children with small-angle Esotropia.
J Optom. 2024 Jul-Sep;17(3):100490. doi: 10.1016/j.optom.2023.100490. Epub 2023 Dec 6.
8
Effect of Fresnel prism in small-angle esotropia (≤ 20 prism diopters) with fixation preference.
Graefes Arch Clin Exp Ophthalmol. 2025 Mar;263(3):857-865. doi: 10.1007/s00417-024-06662-z. Epub 2024 Oct 15.
9
Evaluation of a Novel Virtual Reality Simulated Alternate Cover Test to Assess Strabismus: A Prospective, Masked Study.
Am J Ophthalmol. 2025 Jan;269:266-272. doi: 10.1016/j.ajo.2024.08.042. Epub 2024 Sep 6.
10
Augmented-dose surgery based on the single Maddox rod test for acute acquired comitant esotropia.
J AAPOS. 2024 Dec;28(6):104037. doi: 10.1016/j.jaapos.2024.104037. Epub 2024 Nov 12.

引用本文的文献

2
Evaluation of a Novel Virtual Reality Simulated Alternate Cover Test to Assess Strabismus: A Prospective, Masked Study.
Am J Ophthalmol. 2025 Jan;269:266-272. doi: 10.1016/j.ajo.2024.08.042. Epub 2024 Sep 6.
4
Prevalence and clinical characteristics of age-related distance esotropia.
Strabismus. 2022 Sep;30(3):115-120. doi: 10.1080/09273972.2022.2097707. Epub 2022 Jul 7.
5
COMPUTER BASED EYE TRACKER FOR DETECTION OF MANIFEST STRABISMUS.
Acta Clin Croat. 2021 Dec;60(4):683-694. doi: 10.20471/acc.2021.60.04.16.
7
Interventions for visual field defects in people with stroke.
Cochrane Database Syst Rev. 2019 May 23;5(5):CD008388. doi: 10.1002/14651858.CD008388.pub3.
8
Infrared photographs with a selective wavelength filter to diagnose small-angle esotropia in young children.
Graefes Arch Clin Exp Ophthalmol. 2019 Mar;257(3):645-650. doi: 10.1007/s00417-019-04240-2. Epub 2019 Jan 19.
9
Use of a Binocular Optical Coherence Tomography System to Evaluate Strabismus in Primary Position.
JAMA Ophthalmol. 2018 Jul 1;136(7):811-817. doi: 10.1001/jamaophthalmol.2018.1973.
10
A quantitative analysis method for comitant exotropia using video-oculography with alternate cover.
BMC Ophthalmol. 2018 Mar 22;18(1):80. doi: 10.1186/s12886-018-0747-9.

本文引用的文献

1
Defining real change in prism-cover test measurements.
Am J Ophthalmol. 2008 Feb;145(2):381-5. doi: 10.1016/j.ajo.2007.09.012. Epub 2007 Nov 28.
4
Inter-examiner repeatability of heterophoria tests.
Optom Vis Sci. 1998 Oct;75(10):719-26. doi: 10.1097/00006324-199810000-00016.
6
Ophthalmic prisms. Measurement errors and how to minimize them.
Ophthalmology. 1983 Mar;90(3):204-10. doi: 10.1016/s0161-6420(83)34572-3.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验