LUX-Dx 可插入式心脏监测器远程编程和性能(LUX-Dx PERFORM)研究的初步结果。

Preliminary results from the LUX-Dx insertable cardiac monitor remote programming and performance (LUX-Dx PERFORM) study.

机构信息

Cardiac Rythm Management, Boston Scientific, St Paul, Minnesota, USA.

Cardiac Arrhythmia Service, Florida Atlantic University CESCOM, Boca Raton, Florida, USA.

出版信息

Clin Cardiol. 2023 Jan;46(1):100-107. doi: 10.1002/clc.23930. Epub 2022 Oct 8.

Abstract

Despite the wide adoption of insertable cardiac monitors (ICMs), high false-positive rates, suboptimal signal quality, limited ability to detect atrial flutter, and lack of remote programming remain challenging. The LUX-Dx PERFORM study was designed to evaluate novel technologies engineered to address these issues. Here, we present preliminary results from the trial focusing on the safety of ICM insertion, remote monitoring rates, and the feasibility of remote programming. LUX-Dx PERFORM is a multicenter, prospective, single-arm, post-market, observational study with planned enrollment of up to 827 patients from 35 sites in North America. A preliminary cohort consisting of the first 369 patients who were enrolled between March and October 2021 was selected for analysis. Three hundred sixty-three (363) patients had ICM insertions across inpatient and outpatient settings. The mean time followed was 103.4 ± 61.8 days per patient. The total infection rate was 0.8% (3/363). Interim results show high levels of remote monitoring with a median 94% of days with data transmission (interquartile range: 82-99). Thirteen (13) in-clinic and 24 remote programming sessions were reported in 34 subjects. Reprogramming examples are presented to highlight signal quality, the ability to detect atrial flutter, and the positive impact of remote programming on patient management. Interim results from LUX-Dx PERFORM study demonstrate the safety of insertion, high data transmission rates, the ability to detect atrial flutter, and the feasibility of remote programming to optimize arrhythmia detection and improve clinical workflow. Future results from LUX-Dx PERFORM will further characterize improvements in signal quality and arrhythmia detection.

摘要

尽管可植入式心脏监测器(ICM)已广泛应用,但仍存在高假阳性率、信号质量不佳、检测房扑能力有限以及缺乏远程编程等挑战。LUX-Dx PERFORM 研究旨在评估旨在解决这些问题的新技术。在这里,我们介绍了该试验的初步结果,重点介绍了 ICM 插入的安全性、远程监测率以及远程编程的可行性。LUX-Dx PERFORM 是一项多中心、前瞻性、单臂、上市后、观察性研究,计划在北美 35 个地点招募多达 827 名患者。选择初步队列(由 2021 年 3 月至 10 月期间招募的前 369 名患者组成)进行分析。363 名患者在住院和门诊环境中进行了 ICM 插入。每位患者的平均随访时间为 103.4±61.8 天。总感染率为 0.8%(3/363)。中期结果显示远程监测水平较高,中位数 94%的天数有数据传输(四分位距:82-99)。在 34 名患者中报告了 13 次门诊和 24 次远程编程。呈现了重新编程的示例,以突出信号质量、检测房扑的能力以及远程编程对患者管理的积极影响。LUX-Dx PERFORM 研究的中期结果表明插入的安全性、高数据传输率、检测房扑的能力以及远程编程优化心律失常检测和改善临床工作流程的可行性。LUX-Dx PERFORM 的未来结果将进一步描述信号质量和心律失常检测的改善情况。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/4b88/9849434/b5ae978f726d/CLC-46-100-g002.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验