Biosensors (Basel). 2022 Sep 20;12(10):771. doi: 10.3390/bios12100771.
Procalcitonin (PCT) is considered a sepsis and infection biomarker. Herein, an interdigitated electrochemical immunosensor for the determination of PCT has been developed. The interdigitated electrode was made of the laser-engraved graphene electrode decorated with gold (LEGE/Au). The scanning electron microscopy indicated the LEGE/Au has been fabricated successfully. After that, the anti-PTC antibodies were immobilized on the surface of the electrode by using 3-mercaptopropionic acid. The electrochemical performance of the fabricated immunosensor was studied using electrochemical impedance spectroscopy (EIS). The EIS method was used for the determination of PCT in the concentration range of 2.5-800 pg/mL with a limit of detection of 0.36 pg/mL. The effect of several interfering agents such as the C reactive protein (CRP), immunoglobulin G (IgG), and human serum albumin (HSA) was also studied. The fabricated immunosensor had a good selectivity to the PCT. The stability of the immunosensor was also studied for 1 month. The relative standard deviation (RSD) was obtained to be 5.2%.
降钙素原 (PCT) 被认为是脓毒症和感染的生物标志物。在此,我们开发了一种用于测定 PCT 的叉指式电化学免疫传感器。叉指电极由激光刻蚀石墨烯电极修饰金(LEGE/Au)制成。扫描电子显微镜表明已成功制备了 LEGE/Au。之后,通过使用 3-巯基丙酸将抗-PTC 抗体固定在电极表面上。使用电化学阻抗谱(EIS)研究了所制备的免疫传感器的电化学性能。EIS 方法用于在 2.5-800 pg/mL 的浓度范围内测定 PCT,检测限为 0.36 pg/mL。还研究了几种干扰物,如 C 反应蛋白(CRP)、免疫球蛋白 G(IgG)和人血清白蛋白(HSA)的影响。所制备的免疫传感器对 PCT 具有良好的选择性。还研究了免疫传感器的稳定性,在 1 个月内。得到的相对标准偏差(RSD)为 5.2%。