This study aims to evaluate the effect of different air computed tomography (CT) numbers of the image value density table (IVDT) on the retrospective dose calculation of head-and-neck (HN) radiotherapy using TomoTherapy megavoltage CT (MVCT) images. The CT numbers of the inside and outside air and each tissue-equivalent plug of the "Cheese" phantom were obtained from TomoTherapy MVCT. Two IVDTs with different air CT numbers were created and applied to MVCT images of the HN anthropomorphic phantom and recalculated by Planned Adaptive to verify dose distribution. We defined the recalculation dose with MVCT images using both inside and outside air of the IVDT as and , respectively. Treatment planning doses calculated on kVCT images were compared with those calculated on MVCT images using two different IVDT tables, namely, and . The difference between average MVCT numbers ±1 standard deviation on inside and outside air of the calibration phantom was 65 ± 36 HU. This difference in MVCT number of air exceeded the recommendation lung tolerance for dose calculation error of 2%. The dose differences between the planning target volume (PTV): D , D , D and the organ at risk (OAR): D , D recalculated by and using MVCT images were a maximum of 0.7% and 1.2%, respectively. Recalculated doses to the PTV and OAR with MVCT showed that was 0.5%-0.7% closer to the kVCT treatment planning dose than . This study showed that was more accurate than in recalculating the dose HN cases of MVCT using TomoTherapy.
本研究旨在评估不同空气 CT 数图像值密度表 (IVDT) 对使用 TomoTherapy 兆伏 CT (MVCT) 图像进行头颈部 (HN) 放疗的回溯剂量计算的影响。通过 TomoTherapy MVCT 获得“奶酪”模体内部和外部空气以及每个组织等效插件的 CT 数。创建了两个具有不同空气 CT 数的 IVDT,并将其应用于 HN 人体模型的 MVCT 图像,并通过 Planned Adaptive 重新计算以验证剂量分布。我们将使用 IVDT 内部和外部空气重新计算的 MVCT 图像的剂量定义为 和 ,分别。在 kVCT 图像上计算的治疗计划剂量与在使用两个不同 IVDT 表( 和 )的 MVCT 图像上计算的剂量进行比较。校准模体内部和外部空气的平均 MVCT 数的差异为 65 ± 36 HU。空气的 MVCT 数的这种差异超过了推荐的肺剂量计算误差耐受度 2%。计划靶区 (PTV):D 、D 、D 和危及器官 (OAR):D 、D ,使用 MVCT 图像的 和 重新计算的剂量差异最大为 0.7%和 1.2%。MVCT 重新计算的 PTV 和 OAR 剂量显示, 比 更接近 kVCT 治疗计划剂量,差异在 0.5%-0.7%之间。本研究表明,在使用 TomoTherapy 对 MVCT 的 HN 病例进行剂量重算时, 比 更准确。