Abstract

BACKGROUND

Phacoemulsification has been the mainstay method for extracapsular cataract extraction surgery in the anterior segment; for cases of posterior drop of lens fragments into the vitreous, a posterior segment phacoemulsification instrument (fragmatome; Alcon, Inc., Fort Worth, TX) can be employed to remove the dislocated lens materials. Studies have reported on thermal injury to the cornea during phacmoemulsification of the anterior segment. However, few studies have investigated thermal burn in the simultaneous sclera and eyelid induced by the fragmatome. Currently, there is no reports and lack of optimal strategy for the management of nucleus drop in a vitreous cavity filled with silicon oil.

CASE PRESENTATION

We present the case of a 53-year-old male patient with a thermal burn wound on the upper eyelid and sclera following phacoemulsification for a dropped lens in a silicone oil-filled vitreous. We further designed an experiment to verify our hypothesis that thermal injury could be induced by the high temperature of the metal tip during phacoemulsification in silicone oil. In our experiment, during 420 s of continuous ultrasonic wave, the temperature of the fragmatome tip in the balanced salt solution (BSS) increased from 22.0 to 24.0 ºC, while the temperature of the fragmatome tip in the silicone oil group increased from 22.0 to 43.0 ºC.

CONCLUSIONS

The temperature of the fragmatome tip increased significantly in silicone oil compared to BSS in the experiment. Thus, physicians should be aware of possible thermal complications when using fragmatome in eyes filled with silicone oil.

摘要

背景

超声乳化白内障吸除术已成为眼前节白内障囊外摘除术的主要方法;对于晶状体后段碎片落入玻璃体的情况,可以使用后节超声乳化仪(粉碎器;爱尔康公司,德克萨斯州沃思堡)来去除脱位的晶状体物质。研究报道了在前节超声乳化过程中角膜热损伤。然而,很少有研究报道粉碎器同时引起巩膜和眼睑的热烧伤。目前,对于充满硅油的玻璃体中核块下降,尚无报道和缺乏最佳处理策略。

病例介绍

我们报告了 1 例 53 岁男性患者,在硅油填充的玻璃体中进行脱位晶状体超声乳化术后,出现上眼睑和巩膜热烧伤。我们进一步设计了一项实验来验证我们的假设,即在硅油中进行超声乳化时,金属尖端的高温会导致热损伤。在我们的实验中,在 420 秒的连续超声波作用下,平衡盐溶液(BSS)中粉碎器尖端的温度从 22.0 升至 24.0°C,而硅油组粉碎器尖端的温度从 22.0 升至 43.0°C。

结论

实验中,硅油中粉碎器尖端的温度明显高于 BSS。因此,当在充满硅油的眼睛中使用粉碎器时,医生应该意识到可能出现热并发症。