Abstract

BACKGROUND

Several variations of the combined thrombectomy technique for acute ischemic stroke using a stent retriever and aspiration catheter have been described. The aim of our study was to assess how the presence of the microcatheter and stent retriever affect the basic aspiration parameters, namely, flow rate and aspiration force, depending on their position within the aspiration catheter.

METHODS

Two experimental set-ups were designed to assess changes in flow rate and aspiration force according to the position of the stent retriever and microcatheter within the aspiration catheter.

RESULTS

The transition of the stent retriever and microcatheter from the distal to proximal position resulted in a progressive increase in the flow rate, but with no impact on aspiration force. Additionally, the size of the stent retriever had no significant effect on flow rate changes and the reduction in flow rate was related to the microcatheter diameter. Negative pressure generated inside the aspiration catheter impacted on its distal segment located beyond the radiopaque marker, thus leading to its partial collapse. As a consequence, the measured aspiration force was lower than the theoretical aspiration force level for all tested aspiration catheters.

CONCLUSIONS

In our experimental model, the position of the stent retriever and microcatheter within the aspirator catheter affected the flow rate, but not the aspiration force. Negative pressure generated within the aspiration catheter appeared to determine a partial collapse of the distal segment that resulted in a less effective aspiration force than the theoretical aspiration force level.

摘要

背景

已经描述了几种使用支架取栓器和抽吸导管的联合血栓切除术技术的变体,用于急性缺血性脑卒中。我们的研究目的是评估支架取栓器和微导管的存在如何根据其在抽吸导管内的位置影响基本抽吸参数,即流速和抽吸力。

方法

设计了两种实验装置,以根据支架取栓器和微导管在抽吸导管内的位置评估流速和抽吸力的变化。

结果

支架取栓器和微导管从远端向近端的过渡导致流速逐渐增加,但对抽吸力没有影响。此外,支架取栓器的尺寸对流速变化没有显著影响,流速的降低与微导管的直径有关。抽吸导管内产生的负压会影响位于不透射线标记物远端的导管的远端部分,从而导致其部分塌陷。因此,对于所有测试的抽吸导管,测量的抽吸力均低于理论抽吸力水平。

结论

在我们的实验模型中,支架取栓器和微导管在抽吸导管内的位置会影响流速,但不会影响抽吸力。抽吸导管内产生的负压似乎导致远端部分部分塌陷,导致抽吸力低于理论抽吸力水平。