Abstract

BACKGROUND AND PURPOSE

Thrombectomy in distal, medium vessels is a topic of increasing interest. To date, there are few studies focused on performance of ≤5F catheters in medium vessels. The purpose of this study is to compare the performance of the 3F, 4F, and 5F MIVI Neuroscience Q Catheters versus Penumbra 3F, 4F, and MicroVention Sofia 5F Catheters.

METHODS

Using methods, we assessed and compared the following parameters: aspiration flow rates, clot uncorking forces, impulse, and clot ingestion. For flow rate, each aspiration catheter was immersed in a cylindrical container. Flow rate at one second was used to calculate impulse. For clot uncorking force, the force required to disengage a catheter from a simulated clot was recorded. For ingestion, we measured time to ingest soft and medium stiffness synthetic clots.

RESULTS

The measured flow rates without a stent retriever for the Q3, Q4, and Q5 catheters were 3.54 ml/s, 5.32 ml/s, and 6.87 ml/s. The measured flow rates without a stent retriever for the 3MAX, 4MAX, and 5F Sofia were 1.46 ml/s, 2.56 ml/s, and 1.73 ml/s. The impulse calculated for one second was 26 mNs for Q5 vs 9 mNs for Sofia 5, 35 mNs for Q4 vs 15 mNs for 4Max< and 35 mNs for Q3 vs 9 mNs for 3Max. The average system ingestion for Q was significantly faster than the competitive catheters.

CONCLUSIONS

The Q catheters demonstrated higher flow rates, higher uncorking force, and faster complete clot ingestion than competitive catheters.

摘要

背景与目的

远端中等血管血栓切除术是一个越来越受关注的话题。迄今为止,很少有研究关注≤5F导管在中等血管中的性能。本研究的目的是比较3F、4F和5F MIVI神经科学Q导管与Penumbra 3F、4F和MicroVention Sofia 5F导管的性能。

方法

我们采用以下方法评估和比较了以下参数:抽吸流速、血栓解堵力、冲量和血栓摄取。对于流速,将每个抽吸导管浸入一个圆柱形容器中。用一秒钟时的流速来计算冲量。对于血栓解堵力,记录将导管从模拟血栓中分离所需的力。对于摄取,我们测量了摄取软质和中等硬度合成血栓的时间。

结果

Q3、Q4和Q5导管在没有支架回收器的情况下测得的流速分别为3.54 ml/s、5.32 ml/s和6.87 ml/s。3MAX、4MAX和5F Sofia在没有支架回收器的情况下测得的流速分别为1.46 ml/s、2.56 ml/s和1.73 ml/s。Q5导管一秒钟计算出的冲量为26 mNs,而Sofia 5导管为9 mNs;Q4导管为35 mNs,4Max导管为15 mNs;Q3导管为35 mNs,3Max导管为9 mNs。Q导管的平均系统摄取速度明显快于竞争导管。

结论

与竞争导管相比,Q导管显示出更高的流速、更高的解堵力和更快的完全血栓摄取速度。