Technical Institute of Babylon, Al-Furat Al-Awsat Technical University (ATU), Kufa 54003, Iraq.
Center for Life Science Automation (celisca), University of Rostock, 18119 Rostock, Germany.
Sensors (Basel). 2022 Oct 25;22(21):8161. doi: 10.3390/s22218161.
Air pollution and leakages of hazardous and toxic gases and chemicals are among the dangers that frequently occur at automated chemical and life science laboratories. This type of accident needs to be processed as soon as possible to avoid the harmful side effects that can happen when a human is exposed. Nitrogen oxides and volatile organic compounds are among the most prominent indoor air pollutants, which greatly affect the lifestyles in these places. In this study, a commercial MOX gas sensor, SGP41, was embedded in an IoT environmental sensor node for hazardous gas detection and alarm. The sensor can detect several parameters, including nitrogen oxide index (NOx-Index) and volatile organic compound index (VOC-Index). Several tests were conducted to detect the leakage of nitrogen oxides and volatile organic compounds in different concentrations and volumes, as well as from different leakage distances, to measure the effect of these factors on the response speed and recovery time of the sensors used. These factors were also compared between the different sensors built into the sensor node to give a comprehensive picture of the system used. The system testing results revealed that the SGP41 sensor is capable of implementing the design purposes for the target parameters, can detect a small NO gas leakage starting from 0.3% volume, and can detect all the tested VOC solvents ≥ 100 µL.
空气污染以及危险和有毒气体及化学品的泄漏是自动化化学和生命科学实验室中经常发生的危险之一。此类事故需要尽快处理,以避免人体暴露时可能产生的有害副作用。氮氧化物和挥发性有机化合物是最主要的室内空气污染物之一,它们极大地影响了这些场所的生活方式。在本研究中,商业 MOX 气体传感器 SGP41 被嵌入物联网环境传感器节点中,用于危险气体检测和报警。该传感器可以检测多个参数,包括氮氧化物指数(NOx-Index)和挥发性有机化合物指数(VOC-Index)。进行了多次测试,以检测不同浓度和体积的氮氧化物和挥发性有机化合物的泄漏,以及从不同泄漏距离的泄漏,以测量这些因素对所使用传感器的响应速度和恢复时间的影响。还比较了传感器节点中内置的不同传感器之间的这些因素,以全面了解所使用的系统。系统测试结果表明,SGP41 传感器能够实现目标参数的设计目的,可以检测从 0.3%体积开始的小 NO 气体泄漏,并且可以检测所有测试的 VOC 溶剂≥100µL。