Diopsys NOVA fixed-luminance flicker full-field electroretinogram (ffERG) device is a potential adjunct to conventional flicker ffERG testing for assessing cone cell function. Magnitude of measured electrical response is known to vary with pupil size in conventional ffERG testing. The index study characterizes the relationship between magnitude of measured electrical activity and pupil size, both pupil diameter and pupil area, for this device.
Seventeen patients (34 eyes) with no known ocular diseases were enrolled in the study. Electrophysiologic function of cone cells was evaluated using fixed-luminance flicker ffERG before and after dilation. Linear regression models, with inter-eye correlations controlled as fixed-effects, were used to characterize the effect of pupil dilation on the magnitude of the measured responses.
Mean age of study patients was 33.5 (standard deviation 7.4 years), and 35.3% of the subjects were female. Mean value of electrical response magnitude was 10.07±2.79µV before dilation and 15.30±4.08µV after dilation. The correlations of ERG magnitude with pupil diameter and with pupil area were not significant for either dilated or undilated eyes considered separately but were highly significant (p<0.001) for dilated and undilated eyes considered in aggregate. ERG magnitude tended to increase by 1.08 µV for every 1 mm increase in pupillary diameter.
An increase in pupil size, both pupil diameter and pupil area, is significantly associated with an increase in flicker ffERG magnitude recorded by the Diopsys device, suggesting that pupil size should be measured and considered when making clinical judgments based on the flicker ffERGs recorded by the device, and that pupil size-specific reference ranges could improve the clinical utility of the device.
Diopsys NOVA固定亮度闪烁全视野视网膜电图(ffERG)设备是评估锥体细胞功能的传统闪烁ffERG测试的一种潜在辅助手段。在传统的ffERG测试中,已知测量的电反应幅度会随瞳孔大小而变化。本指标研究描述了该设备测量的电活动幅度与瞳孔大小(包括瞳孔直径和瞳孔面积)之间的关系。
17例(34只眼)无已知眼部疾病的患者纳入本研究。在散瞳前后使用固定亮度闪烁ffERG评估锥体细胞的电生理功能。采用线性回归模型,将眼间相关性作为固定效应进行控制,以描述瞳孔扩张对测量反应幅度的影响。
研究患者的平均年龄为33.5岁(标准差7.4岁),35.3%的受试者为女性。散瞳前电反应幅度的平均值为10.07±2.79µV,散瞳后为15.30±4.08µV。单独考虑散瞳或未散瞳的眼睛时,ERG幅度与瞳孔直径和瞳孔面积的相关性均不显著,但综合考虑散瞳和未散瞳的眼睛时,相关性非常显著(p<0.001)。瞳孔直径每增加1mm,ERG幅度往往增加1.08µV。
瞳孔大小(包括瞳孔直径和瞳孔面积)的增加与Diopsys设备记录的闪烁ffERG幅度增加显著相关,这表明在基于该设备记录的闪烁ffERG进行临床判断时,应测量并考虑瞳孔大小,并且特定瞳孔大小的参考范围可以提高该设备的临床实用性。