长期使用enFlow输液加温器时可浸出元素的水平。

Levels of leachable elements from long-term use of enFlow fluid warmer.

作者信息

Bayoro Danielle, Groepenhoff Herman, Pedro Michael, Rose Edward A, Waldmann Andreas D

机构信息

Medical Affairs, Vyaire Medical, Mettawa, IL, USA.

出版信息

SAGE Open Med. 2022 Jul 5;10:20503121221108927. doi: 10.1177/20503121221108927. eCollection 2022.

Abstract

OBJECTIVE

In the delivery of intravenous fluids, in-line warming devices frequently transfer heat using a metal heating plate, which if uncoated can risk elution. This bench study examined extractable elements detected following long-term use of the parylene-coated enFlow Disposable IV/Blood Warmer.

METHODS

We tested 16 clinically relevant challenge fluids typical of the surgical setting, including commercially available single donor blood and blood products as well as intravenous saline and electrolyte solutions. After 72 h of warming at 40°C (104°F) via the enFlow, analytical chemistry identified and quantified the most clinically significant extractable elements (arsenic, barium, cadmium, copper, and lead) to estimate chemical exposure. We also measured the extracted concentrations of these five elements following simulated use of the device with three solutions (Sterofundin ISO, Plasma-Lyte 148, and whole blood) that were pumped through the warmed device at two different flow rates (0.2 and 5.5 mL min).

RESULTS

Across all scenarios of acute and long-term exposures for different populations, the enFlow demonstrated low toxicological risks as measured by the calculation of tolerable exposure for extracted arsenic, barium, cadmium, copper, and lead.

CONCLUSION

The results suggest biological safety for the use of parylene-coated enFlow with a variety of intravenous solutions and in different therapeutic scenarios.

摘要

目的

在静脉输液过程中,在线加温装置通常使用金属加热板来传递热量,如果未涂层则存在洗脱风险。这项实验台研究检测了聚对二甲苯涂层的enFlow一次性静脉输液/血液加温器长期使用后检测到的可提取物元素。

方法

我们测试了16种手术环境中典型的临床相关挑战性液体,包括市售的单供体血液和血液制品以及静脉生理盐水和电解质溶液。通过enFlow在40°C(104°F)下加温72小时后,分析化学方法鉴定并定量了临床上最重要的可提取物元素(砷、钡、镉、铜和铅),以估计化学暴露量。我们还测量了在使用该装置模拟三种溶液(Sterofundin ISO、Plasma-Lyte 148和全血)后这五种元素的提取浓度,这三种溶液以两种不同流速(0.2和5.5 mL/min)泵入加温装置。

结果

在针对不同人群的所有急性和长期暴露场景中,通过计算提取的砷、钡、镉、铜和铅的可耐受暴露量来衡量,enFlow显示出较低的毒理学风险。

结论

结果表明,在各种静脉输液溶液和不同治疗场景中使用聚对二甲苯涂层的enFlow具有生物安全性。