Ertel Monica K, Seibold Leonard K, Patnaik Jennifer L, Kahook Malik Y
Department of Ophthalmology, University of Colorado, UC Health Sue Anschutz-Rodgers Eye Center, Aurora, Colorado 80045, USA.
Int J Ophthalmol. 2022 Dec 18;15(12):2022-2027. doi: 10.18240/ijo.2022.12.19. eCollection 2022.
To compare intraocular pressure (IOP) readings obtained with Perkins tonometry, iCare Home, iCare 200, and Tonopen to IOP readings obtained with the manometer of a perfusion system to assess the accuracy and reproducibility of each method of tonometry at set pressures.
The IOP of human cadaveric eyes (=2) was measured using a manometer inserted into the eye through the optic nerve. IOP measurements were obtained using a Perkins tonometer, iCare Home, iCare 200, and Tonopen. These measurements were compared to set point IOP measurements of a manometer to determine accuracy and reproducibility of each device.
Mean IOP readings obtained with the Perkins tonometer compared to manometer readings demonstrated a difference of -1.0±5.0 mm Hg (=0.45), indicating a lower reading on average than manometery although not significant. Mean IOP difference between iCare 200 and manometer was 5.3±2.2 mm Hg (<0.0001). Mean difference in IOP between iCare Home and manometer was 3.5±2.4 mm Hg (=0.0004). Mean IOP difference compared to manometer was 4.6±4.0 mm Hg for the Tonopen (<0.0001). IOP measurements obtained with the Perkins tonometer demonstrated a standard deviation of 5.0 mm Hg while the Tonopen measurements demonstrated a 4.0 mm Hg standard deviation. In comparison, iCare 200 and iCare Home demonstrated 2.2 and 2.4 mm Hg standard deviation, respectively.
Applanation tonometry produces more accurate IOP readings than rebound tonometry or Tonopen, however it demonstrates greater variability than the other forms of tonometry. Rebound tonometry is more reproducible but tends to over-estimate IOP.
比较使用珀金氏眼压计、iCare Home、iCare 200和非接触眼压计(Tonopen)所测得的眼压读数与使用灌注系统压力计所测得的眼压读数,以评估在设定压力下每种眼压测量方法的准确性和可重复性。
通过将压力计经视神经插入人尸体眼内来测量眼压(n = 2)。使用珀金氏眼压计、iCare Home、iCare 200和非接触眼压计(Tonopen)进行眼压测量。将这些测量结果与压力计的设定点眼压测量结果进行比较,以确定每种设备的准确性和可重复性。
与压力计读数相比,使用珀金氏眼压计获得的平均眼压读数差异为-1.0±5.0 mmHg(n = 0.45),表明平均读数虽低于压力计测量值,但差异无统计学意义。iCare 200与压力计之间的平均眼压差异为5.3±2.2 mmHg(P<0.0001)。iCare Home与压力计之间的眼压平均差异为3.5±2.4 mmHg(P = 0.0004)。非接触眼压计(Tonopen)与压力计相比的眼压平均差异为4.6±4.0 mmHg(P<0.0001)。使用珀金氏眼压计获得的眼压测量值的标准差为5.0 mmHg,而非接触眼压计(Tonopen)测量值的标准差为4.0 mmHg。相比之下,iCare 200和iCare Home的标准差分别为2.2和2.4 mmHg。
压平眼压计比回弹眼压计或非接触眼压计(Tonopen)产生更准确的眼压读数,然而,它比其他形式的眼压测量法表现出更大的变异性。回弹眼压计更具可重复性,但往往高估眼压。